溴化镧探测器的晶体表征与全能峰效率计算 被引量:4

Size adjustment and calculation of full-energy peak efficiency for the LaBr_3 scintillation detector
下载PDF
导出
摘要 无源效率刻度技术已在辐射探测器的效率刻度中得到了广泛的应用。本文基于点源效率实验及蒙特卡罗模拟计算方法,对溴化镧探测器的晶体尺寸进行了调整;在此基础上,计算得到不同位置处全能峰效率的计算值。点源效率实验的验证结果表明,在121~1 332 keV能量范围内,γ光子全能峰效率的计算值与实验值的相对偏差小于±5%。 The technique of virtual calibration has been widely applied to field of calibration of the radiation detector. In this paper,the dimensions of the LaBr_3 detector were adjusted based on the pointsource experiments with Monte Carlo method. A series of full-energy peak efficiency of the detector were then simulated in different positions. The results of the validation experiment showed that the relative deviations between calculated efficiencies and measurement values are within ± 5% in the energy range of 121—1 332 keV.
作者 曹勤剑 夏三强 刘立业 肖运实 汪屿 李华 Cao Qinjian Xia Sanqiang Liu Liye Xiao Yunshi Wang Yu Li Hua(China Institute of Atomic Energy, Beijing 102413 China Institute for Radiation Protection, Taiyuan 030006)
出处 《辐射防护》 CAS CSCD 北大核心 2016年第6期364-367,416,共5页 Radiation Protection
基金 国家自然科学基金资助项目(项目编号:11305146 11275174)
关键词 溴化镧探测器 蒙特卡罗 晶体调整 效率刻度 LaBr3 scintillation detector Monte Carlo method crystal adjustment efficiency calibration
  • 相关文献

参考文献4

二级参考文献36

  • 1牟海维,韩建,李中伟.一种新型的时间-幅度变换器[J].核电子学与探测技术,2004,24(6):612-615. 被引量:3
  • 2谢建军,杨培志,廖晶莹.卤化镧系LnX_3(Ce)闪烁晶体的研究进展[J].无机材料学报,2005,20(3):522-528. 被引量:18
  • 3张斌全,马吉增,程建平,刘立业,毛永.蒙特卡罗方法计算用于低能光子测量的高纯锗探测器的效率[J].核电子学与探测技术,2005,25(3):274-277. 被引量:26
  • 4Alonso M,et al. Monte Carlo estimation of absorbed dose to organs in computed tomography[J].Health Phys,2002,82(2):233. 被引量:1
  • 5Stabin MG,et al.Photon specific absorbed fractions culculated in the trunk of an adult male voxel-based phantom[J]. Health Phys,2002,82(1):21. 被引量:1
  • 6Borisov N,. et al.A new graphical user interface for fast construction of computation phantoms and mcnp calculations:application to calibration of in vivo measurement systems[J]. Health Phys,2002,83(2):272. 被引量:1
  • 7Bozkurt A, et al Fluence-to-dose conversion coefficients from monoenergetic neutrons below 20 MeV based on the VIP-Man anatomical model[J]. Phys Med Biol,2000,(45):3059. 被引量:1
  • 8Moens L, et al Calculation of the absolute peak efficiency of gamma-ray detectors for different counting geometries[J]. Nucl Instrum Methods,1981,187:451. 被引量:1
  • 9Moens L, et al Calculation of the peak efficiency of high-purity Germanium detectors[J]. Int J Appl Radiat Isot,34(8):1085. 被引量:1
  • 10Hannele Aaltonen, et al Validation of a method for computer calculation of germanium detector efficiencies[J]. Nucl Instr.um Methods ,1994,339: 87. 被引量:1

共引文献42

同被引文献20

引证文献4

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部