期刊文献+

原子荧光光度计测定硒的检出限和精密度 被引量:2

Detection Limit and Precision of Selenium Detected by Atomic Fluorescence Spectrometry
下载PDF
导出
摘要 [目的]验证原子荧光法测定硒的检出限和精密度的准确性。[方法]运用双道原子荧光光度计AFS-9730测定硒的检出限和精密度,检验其是否满足环境保护部新颁布的《水质汞、砷、硒、铋和锑的测定原子荧光法》(HJ 694—2014)对检出限和精密度的要求。[结果]原子分光光度计测定硒的检出限为0.005 2μg/L,相对标准偏差为0.78%。[结论]原子分光光度计测定硒的检出限和精密度满足环境保护部新颁布的《水质汞、砷、硒、铋和锑的测定原子荧光法》(HJ 694—2014)的要求。 [Objective] To research the detection limit and precision of selenium detected by atomic fluorescence spectrometry.[Method] Double channel atomic fluorescence spectrometry AFS-9730 was used to detect the detection limit and precision of selenium.We researched whether the detection limit and precision met the requirements of Water Quality-Determination of Mercury, Arsenic, Selenium, Bismuth and An-timony-Atomic Fluorescence Spectrometry (HJ694-2014).[Result] The detection limit of selenium was 0.005 2 ug/L by atomic fluorescence spectrometry, with the relative standard deviation being 0.78%.[Conclusion] Detecting the precision and detection limit by atomic fluores-cence spectrometry meets the requirements of Water Quality-Determination of Mercury, Arsenic, Selenium, Bismuth and Antimony-Atomic Fluo-rescence Spectrometry (HJ694-2014).
作者 李松涛
出处 《安徽农业科学》 CAS 2016年第29期30-30,34,共2页 Journal of Anhui Agricultural Sciences
关键词 原子荧光光度计 检出限 精密度 Atomic fluorescence spectrometry Selenium Detection limit Precision