MLC类型不同的加速器对鼻咽癌调强放疗计划影响的研究 被引量:9

Effects of types of accelerators of MLC on IMRT plans of nasopharyngeal carcinoma
下载PDF
导出
摘要 目的 :比较配置宽度不同的多叶光栅叶片的加速器对鼻咽癌调强放疗计划的影响,探讨最适合鼻咽癌调强放疗的加速器。方法:选择32例鼻咽癌患者进行调强放疗,利用Pinnacle3 V9.2计划系统分别选择Elekta Precise和Varian 6EX 2台加速器进行调强放射治疗(intensity-modulated radiation therapy,IMRT)计划设计,根据剂量体积直方图对靶区剂量以及危及器官进行综合分析,比较2种计划在靶区适形度和剂量均匀性以及危及器官受照剂量方面的差异。结果:在2种加速器上完成的调强放疗计划的靶区平均剂量之间差别无统计学意义(P=0.863);Varian 6EX加速器的靶区适形度指数和剂量均匀性指数均优于Elekta Precise(P<0.001,P=0.04)。二者在脑干、脊髓、晶体、视神经、腮腺等器官的剂量均无显著差异(P>0.05)。结论:选择Varian 6EX直线加速器完成调强放疗计划,有较好的靶区适形度和剂量均匀性,更适合鼻咽癌的调强放射治疗。 Objective To compare the effects of the accelerators with different widths of multi-leaf collimators on nasop-haryngeal carcinoma(NPC) IMRT plans, and to explore the optimum accelerator for NPC IMRT. Methods Totally 32 NPC patients were selected to undergo IMRT. IMRT plans were made using Pinnacle3 V9.2 TPS on the basis of Elekta Precise and Varian 6EX respectively. The target area dose and involved organs were analyzed with dose-volume histogram. The CI and DHI of target as well as dose received by OARs were compared. Results There were no significant differences in the mean dose of IMRT plans using different accelerators(P=0.863), while the CI and DHI of plans made by Varian 6EX were better than that by Elekta Precise(P〈0.001, P=0.04). There were no significant differences found in the terms of dose received by the OARs, such as brainstem, spinal cord, crystal, optic nerve, parotid gland(P〈0.05). Conclusion Varian 6EX with optimum CI and DHI IMRT plans is suitable for NPC radiation therapy.
出处 《医疗卫生装备》 CAS 2016年第8期78-80,共3页 Chinese Medical Equipment Journal
基金 广西壮族自治区卫计委资助项目(Z2014261)
关键词 多叶光栅 调强放射治疗 鼻咽癌 靶区剂量 multi-leaf collimator intensity-modulated radiation therapy nasopharyngeal carcinoma target area dose
  • 相关文献

参考文献12

二级参考文献73

共引文献129

同被引文献74

引证文献9

二级引证文献29

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部