用KV—CBCT探讨两种下肢固定方法对直肠癌IMRT摆位误差影响 被引量:12

Influence of lower limb movements on setup errors in radiotherapy for rectal cancer using kilovoltage cone-beam CT (KV-CBCT)
原文传递
导出
摘要 目的探讨负压真空垫固定双下肢法与自制泡沫脚垫固定双脚踝法固定体位方式在直肠癌IMRT中对摆位误差的影响。方法2014年选取12例直肠癌患者随机均分成A、B组,A组患者负压真空垫固定双下肢法,B组患者采用自制塑料泡沫脚垫固定双侧脚踝法。采集108人次(A、B组各54人次)治疗前后KV—CBCT图像行灰度平移误差配准,成组t检验两组患者在左右方向(x轴)、上下方向(Y轴)、前后方向(z轴)的摆位误差。结果两组患者y轴摆位误差相近[(2.13±0.64)min:(2.61±1.17)mm,P=0.399];x轴摆位误差A组低于B组[(1.51±0.28)mm:(2.70±1.05)mm,p=0.039],A、B组误差率分别7.41%、42.59%;z轴上A组也低于B组[(1.10±0.29)mm:(2.37±0.71)mm,P=0.002],A、B组误差率分别1.85%、35.19%。结论直肠癌放疗摆位采用负压真空垫固定双下肢法有效避免了双下肢左右平移和内外旋转。从而减少摆位误差.比自制泡沫脚垫固定双脚踝法的精确性更高。 Objective To investigate the influence of fixation of both lower limbs with negative pressure vacuum cushion and fixation of both ankles with self-made foam mat on setup errors in radiotherapy for rectal cancer. Methods A total of 12 patients with rectal cancer were enrolled in 2014 and randomly divided into group A (using negative pressure vacuum cushion) and group B (using self-made foam mat).An offline registration analysis was performed for the images of 108 times (A, B group of 54 times) of kilovohage cone-beam CT (CBCT) before and after treatment. Grey scale translation error registration was used, and the results of registration were analyzed. The setup errors in x-axis (left-right direction) , y-axis (cranial-caudal direction), and z-axis ( anterior-posterior direction) were compared between the two groups. Results There was no significant difference in the absolute setup error in the y-axis between the two groups (2.13±0. 64 mm vs.2.61 ± 1.17 mm,P-0. 399), while group A showed significantly lower absolute setup errors in the x-axis and z-axis than group B (x-axis:1.51±0. 28 mm vs. 2.70±1.05 mm,P=0. 039;with an error rate of 7.41% vs. 42. 59% ;z-axis : 1.10±0. 29 mm vs. 2.37±0. 71 mm, P= 0. 002 ; with an error rate of 1.85% vs. 35.19% ). Conclusions In the radiotherapy positioning for rectal cancer, fixation of both lower limbs with negative pressure vacuum cushion effectively avoids the translation and rotation of both lower limbs, reduces absolute setup errors, and has higher accuracy than fixation of both ankles with self-made foam mat.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2016年第9期955-958,共4页 Chinese Journal of Radiation Oncology
关键词 体层摄影术X线计算机 锥形束 千伏级 真空垫固定 泡沫脚垫固定 摆位误差 直肠肿瘤/放射疗法 Tomography, X-ray computed, cone beam, kilovolt Negative pressure vacuum cushion fixation Self-made foam mat fixation Set-up error Rectal neoplasms/radiotherapy
  • 相关文献

参考文献5

二级参考文献36

  • 1于龙珍,翟振宇,沈君姝.体部肿瘤精确放疗摆位误差分析[J].中国医学物理学杂志,2007,24(3):160-162. 被引量:36
  • 2潘建基,潘才住,陈传本,张秀春,柏朋刚.摆位系统误差对鼻咽癌调强放疗剂量的影响[J].中华放射肿瘤学杂志,2007,16(5):394-396. 被引量:43
  • 3Hunt M. The effect of positional uncertainties on the treatment of primary nasopharynx cancer. Med Phys, 1989,16:456 - 464. 被引量:1
  • 4Stroom JC, de Boer JC, Huizenga H, et al. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability. Int J Radiat Oncol Biol Phys, 1999,43:905-919. 被引量:1
  • 5Stroom JC, Heijmen BJ. Geometrical uncertainties, radiotherapy planning margins, and the ICRU - 62 report. Radiother Oncol, 2002,64:75 - 83. 被引量:1
  • 6ICRU Report 50: prescribing, recording, and reporting photon beam therapy. Bethesda:Intemational Commission on Radiation Units and Measurements(ICRU), 1993 : 1-72. 被引量:1
  • 7ICRU Report 62: prescribing, recording, and reporting photon beam therapy (Supplement to ICRU Report 50). Bethesda: International Commission on Radiation Units and Measurements (ICRU), 1999:1-51. 被引量:1
  • 8Butler EB, The BS, Grant WH, et al. Smart Boost: a new accelerated fractionation schedule for the treatment of head and neck cancer with intensity modulated radiotherapy. Int J Radiat Oncol Biol Phys, 1999,45:21-32. 被引量:1
  • 9Kuppersimith RB, Greao SC, The BS, et al. Intensity modulated radiotherapy: first result with this new technology on neoplasms of the head and neck. Ear Nose Throat J, 1999,78:238-241. 被引量:1
  • 10Moham R, Wu Q, Manning M, et al. Radiobiological considerations in the design of fractionation strategies for intensity modulated radiation therapy of head and neck cancers. Int J Radiat Oncol Biol Phys, 2000,46:619-630. 被引量:1

共引文献90

同被引文献98

引证文献12

二级引证文献101

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部