摘要 设计了一种利用推衍十二导联原理,以少电极得到常规十二导联心电信息的新型检测系统。它以手机为主机,采用ADS1294和MSP430F149单片机构成从机,并用蓝牙通信接口进行数据传输。数据处理与显示软件采用Android编程。共模干扰和50Hz工频干扰分别采用数字滤波器和数字陷波器去除。仅用五个电极的实验样机与九个电极的标准十二导联心电图机的对比实验结果表明,两者在诊断参数上的平均误差为3.76%,波形匹配很好。新系统在临床诊断和家庭保健等方面都有较大的应用前景。 In this study,a novel system is proposed for the detection of the EASI-derived 12-lead ECG,which outputs the standard 12-lead ECG with fewer electrodes. It uses a mobile phone as the master unit.Makes the slave with ADS1294 amplifier chip and MSP430F149 micro-controller,and transmits data on the Bluetooth communication interface.Its software for data processing and display is programmed in Android.The common-mode interference and the 50Hz interference are removed by using a digital filter and a digital notch tilter,respectively.
作者 黄奇奇 杨圣
出处 《工业控制计算机》 2016年第6期49-50,52,共3页 Industrial Control Computer
基金 国家自然科学基金项目(61172036)资助
关键词 推衍十二导联 心电采集 ADS1294 ANDROID EASI,ECG acquisition,ADS 1294,Android
  • 相关文献

参考文献11

二级参考文献31

  • 1陈天华,陈茜.一种消除心电信号噪声的数字滤波方法[J].世界科学技术-中医药现代化,2005,7(1):123-126. 被引量:7
  • 2刘玉英,侯岩芳.冠心病患者无症状性心肌缺血心电监护的分析及护理[J].中国实用护理杂志(中旬版),2006,22(3):11-12. 被引量:3
  • 3[1]安刚,薛富善主编 .现代麻醉学技术 .北京:科学技术文献出版社,2001: 526-529 被引量:4
  • 4[1]Wehr G,Peters R J,Khalife K.A vector-based,5-electrode,12-lead monitoring ECG (EASI) is equivalent to conventional 12-lead ECG for diagnosis of acute coronary syndromes.J Electrocardiol,2006.39:22-28. 被引量:1
  • 5[2]Chantad D,Krittayaphong R,Komoltri C.Derived 12-lead electrocardiogram in the assessment of ST-segment deviation and cardiac rhythm.J Electrocardiol,2006,39:7-12. 被引量:1
  • 6[3]Sejersten M,Wagner G S,Pahlm O,et al.Detection of acute ischemia from the EASI.derived 12-lead electrocardiogram and from the 12-lead electrocardiogram acquired in clinical practice.J Electrocardiol,2007,40:120-126. 被引量:1
  • 7[5]Dower G E,Machado H B,Osborne J A.On deriving the electrocardiogram from vectorcardiographic leads.Clin Cardiol,1980,3:87-95. 被引量:1
  • 8[6]Dower G E,Yakush A,Nazzal S B.Deriving the 12-lead electrocardiogram from four (EASI) electrodes.J Electrocardiol,1988,21 (Suppl):S182-S187. 被引量:1
  • 9[7]Feild D Q,Feldman C L,Horacek B M.Improved EASI coefficients:their derivation,values,and performance.J Electrocardiol,2002,35 (Suppl):23-33. 被引量:1
  • 10[8]Mangano D T,Layug E L,Wallace A,et al.Effect of atenolol on mortality and cardiovascular morbidity after noncardiac surgery.Multicenter Study of Perioperative Ischemia Research Group.N Engl J Med,1996,335:1713. 被引量:1