摘要
对一例发生在ICU(重症监护室)病房火灾的起火原因进行调查,认定起火原因为吊塔吊臂电气故障。对照有关医用电气设备安全标准,分析致灾机理,总结提出此类场所防范措施和火灾事故调查方法,为医用电气设备的火灾调查及预防提供借鉴。
The fire investigation of an ICU ward fire was conduc- ted, and the'fire reason was identified to be the electrical fault of the lifting arm. Comparing with the standards of medical electrical equipment, the fire cause was analyzed, and the fire prevention and investigation methods were summarized, which can provide reference for medical electrical equipment fire investigation.
出处
《消防科学与技术》
CAS
北大核心
2016年第5期726-728,共3页
Fire Science and Technology
关键词
电气火灾
医用电气设备
火灾调查
fire
medical electrical equipment
fire investigation