射波刀脊柱追踪精度检测与评估 被引量:6

Detection and evaluation of spine tracking accuracy of CyberKnife
下载PDF
导出
摘要 目的利用头颈部、肺部和仿真人模体检测射波刀影像引导摆位(IGRT)治疗脊柱追踪精度并进行评估。方法利用CT分别扫描装有胶片的头颈部、肺部和仿真人模体,将影像数据传输至Multi Plan治疗计划系统中。设计颈椎、胸椎、腰椎、尾椎、股骨追踪的模体计划,等中心计划,70%剂量曲线给予420 c Gy。模体置于治疗床上,执行模体计划。扫描照射后的胶片,利用端到端测试(E2E)软件分析照射精度。结果颈椎追踪的精度为0.54 mm和0.68 mm,胸椎为0.50 mm和0.65 mm,腰椎、骶椎、髂骨和股骨分别为0.70、1.99、2.21、2.05 mm。结论对于这3种静态模体,颈椎、胸椎和腰椎的追踪精度均达到亚毫米,骶椎、髂骨和股骨精度为(2.00±0.21)mm,脊柱追踪的精度非常高。 Objective To detect and evaluate spine tracking accuracy of CyberKnife in image guided radiation therapy(IGRT) with phantom of head-and-neck, lung and simulated human. Methods CT was used to scan the head-and-neck phantom, lung phantom and simulated human phantom with films, and the image data was transferred to MuhiPlan treatment planning system. The tracking phantom plans of cervical, thoracic, lumbar, caudal and femur were designed, and the 70 % dosage curve was 420 eGy. The phantoms were placed on treatment bed, and the plans carried out by CyberKnife, The irradiated films were scanned, end to end testing(E2E) software was used to analyze irradiation accuracy. Results The tracking accuracy of cervical was 0.54 mm and 0.68 mm, thoracic was 0.50 mm and 0.65 mm, lumbar was 0.70 ram, sacral was 1.99 ram, iliac was 2.21 mm and femoral was 2.05 mm. Conclusion It is demonstrated that in three static phantoms, the tracking accuracy of cervical, thoracic and lumbar are at submillimeter level, and tracking accuracy of sacral, ilium and femoral is (2.00 ± 0.21) mm. The spine tracking accuracy has the highest precision.
机构地区 解放军第
出处 《生物医学工程与临床》 CAS 2016年第1期26-29,共4页 Biomedical Engineering and Clinical Medicine
基金 首都临床特色应用研究与成果推广项目课题(Z151100004015002)
关键词 射波刀 脊柱追踪 精度 检测 CyberKnife spine tracking accuracy detection
  • 相关文献

参考文献13

  • 1Floriano A, Garcia R, Moreno R, et al. Retrospective evaluation of CTV to PTV margins using CyberKnife in patients with thora- cic tumors[J]. J Appl Clin Med Phys, 2014, 15(6): 4825- 4825. 被引量:1
  • 2Keall P. TH-D-BRD--04: A review of present and near-future methods for real-time target tracking and adaptation[J]. Med Ph- ys, 2015, 42(6): 3738-3738. 被引量:1
  • 3Dieterich S, Cavedon C, Chuang CF, et al. Report of the AAPM t- ask group 135: Quality assurance for robotic radiosurgery[J]. M- ed Phys, 2011, 38(6): 2914-2916. 被引量:1
  • 4Jung J, Song SY, Yoon SM, et al. Verification of accuracy of Cy- berKnife tumor-tracking radiation therapy using patient-speci- fic lung phantoms[J]. Int J Radiat Oncol Biol Phys, 2015, 92(4): 745-753. 被引量:1
  • 5Blanck O, Masi L Damme MC, et al. Film-based delivery qual- ity assurance for robotic radiosurgery: Commissioning and vali- dation[J]. Phys Med, 2015, 31(5): 476-483. 被引量:1
  • 6Boschini FJ, Commander CW, Schreiber EC, et al. Comparison of gold seeds, tight embolization coils, and loose embolization c- oils used as fiducial markers for dynamic tracking using Cyber- Knife[J]. Pract Radiat Oncol, 2013, 3(2 Suppl 1): S`22-S22. 被引量:1
  • 7Veronese I, De Martin E, Martinotti AS, et al. Multi-institutional application of failure mode and effects analysis(FMEA) to Cybe- rKnife stereotactic body radiation therapy(SBRT)[J].Radiat On- col, 2015, 10: 132-132. 被引量:1
  • 8Guo Y, Zhuang H, Zhao L, et al. Influence of different image-gu- ided tracking methods upon the local efficacy of CyberKnife tre- atment in lung tumors[J]. Thorac Cancer, 2015, 6(3): 255- 259. 被引量:1
  • 9徐慧军,李玉,张素静,杨晓,张军华.G4射波刀两年质量保证检测结果的回顾与评价[J].中国医学物理学杂志,2013,30(3):4097-4099. 被引量:8
  • 10A1 Karl MA, Mwidu U, Moftah B. Continuous versus step-by- step scanning mode of a novel 3D scanner for CyberKnife measurements[J]. Appl Radiat Isot, 2015, 105: S8-91. 被引量:1

二级参考文献10

  • 1徐慧军,李玉,张素静,探讨建立G4射波刀系统的验收测试程序.医疗卫生装备.2012,10:355-362. 被引量:3
  • 2soia Dietench等.机器人放射外科的质量保证:AAPM工作组135号报告.医学物理.2011.38(6):2924-2926. 被引量:1
  • 3Christos Antypas.Evaggelos Pantelis.对G4射波刀的影像引导机器人立体定向放射外科系统的性能评估.医学物理.生物医学.2008;534697-4718. 被引量:1
  • 4.Pantelis,L.Petrokokkinos,C.Antypas.G4射波刀机器人立体定向放射外科系统的影像引导质量评估.2009JINST4P05009. 被引量:1
  • 5YuC,MainW,TaylorD,KuduvalliG,Apuzzo ML,AdlerJRJr.对射波刀脊柱放疗准确性所作的仿真体模研究.神经外科学.2004;55:1138-49. 被引量:1
  • 6Wouter Wunderink.Alejandra Mendez Romero,Willyde Kmijf等.通过追踪植入肝脏的金标分析立体定向全身框架中腹部加压对肿瘤呼吸运动的降低作用.放射肿瘤学.生物物理.2008;71:907-915. 被引量:1
  • 7Hoogeman M,Prevost JB,Nuyttens J等.射波刀的肿瘤呼吸追踪系统的临床精确性:通过分析系统记录文件进行评估.放射肿瘤学.生物物理.2009;74:297-303. 被引量:1
  • 8A.K.Ho,D.Fu,C.Cotrutz,S.L.Hancock等.研究通过骨骼追踪进行射波刀脊柱放疗的准确性.神经外科学60,ONSl47-156讨论ONSl56(2007). 被引量:1
  • 9徐慧军,李玉,张素静,朴俊杰,杨晓.G4射波刀蒙特卡罗算法的建立与评价[J].中国医学影像技术,2012,28(10):1932-1935. 被引量:8
  • 10徐慧军,李玉,张素静.测量治疗床、SBRT体架及DAVID系统造成的剂量衰减[J].中国医学物理学杂志,2012,29(6):3718-3720. 被引量:4

共引文献7

同被引文献47

引证文献6

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部