钢丝电极心脏穿刺心内膜起搏对活体犬正常搏动心肌的影响

Effect of endocardiac pacing by steel wire electrode cardial puncture on living dog myocardium with normal heart beat
下载PDF
导出
摘要 目的:观察钢丝电极对活体动物正常心跳情况下心脏穿刺心内膜起搏中起搏参数及心脏的影响,并评价其安全性。方法6只杂种犬麻醉及开胸后,使用环状钢丝电极和钩状钢丝电极分别在右心室相应的4个部位行心脏穿刺心内膜起搏(每只犬各接受8次穿刺),测定每个部位每个电极的有效起搏时间、起搏参数和穿刺并发症(出血时间及出血量);最终两种电极各完成24次测定操作,比较两组各测定值的差异。结果两组心脏起搏成功率均为100%;环状钢丝电极组有效起搏心脏操作时间稍短于钩状钢丝电极组,但两组比较差异无统计学意义(s:18.4±2.3比19.6±4.1,P>0.05);环状钢丝电极组起搏参数如操作时间(s:18.4±2.3比19.6±4.1)、起搏阈值(V:2.1±0.2比2.2±0.8)、R波振幅(mV:11.3±3.2比12.6±4.1)、电极阻抗(Ω:674.2±89.7比668.5±101.3)等与钩状钢丝电极组比较差异均无统计学意义(均P>0.05)。环状钢丝电极组插入和拔出电极后出血时间〔插入电极后(min):4.4±2.3比4.5±3.1,拔出电极后(min):4.1±2.2比4.8±2.5〕、出血量〔插入电极后(mL):2.8±2.4比3.2±3.5,拔出电极后(mL):3.3±1.7比3.5±2.6,均P>0.05〕与钩状钢丝电极组比较差异均无统计学意义(均P>0.05)。结论在活体动物正常心跳情况下,环状钢丝电极与钩状钢丝电极同样在心脏穿刺心内膜起搏中操作性能和起搏功能均良好,对心脏未产生明显损伤作用,有助于迅速建立有效的紧急心脏起搏。 Objective To observe the parameters and effect of endocardial pacing by steel wire electrode cardiac puncture on heart with normal beat in living animal, and evaluate its safety.Methods After anaesthesia and thoracotomy in 6 living dogs with normal heart beat, the pericardia were excised. Steel wire electrodes with annular or hook tips were used respectively at right ventricular 4 corresponding spots to perform cardiac puncture endocardiac pacing (each dog experienced 8 times of puncture); the time from puncture to effective pacing, pacing parameter and puncture complication (time and quantity of bleeding) of each electrode at each spot were recorded. Finally, the two types of electrode completed 24 times of manipulation respectively; the data collected of the two types were compared. Results The cardiac pacing successful rates in the two groups were 100%; the time taken from the beginning of heart puncture to effective pacing in annular tip group was less than that in hook tip group, but the time difference between the two groups showed no statistical significance (s: 18.4±2.3 vs. 19.6±4.1,P 〉 0.05). The parameters of pacing in the annular tip group, such as operation time (s: 18.4±2.3 vs. 19.6±4.1), the threshold value of pacing (V: 2.1±0.2 vs. 2.2±0.8), the amplitude of R wave sensed (mV: 11.3±3.2 vs. 12.6±4.1) and the impedance of electrode (Ω: 674.2±89.7 vs. 668.5±101.3) were not significantly different compared with those in the hook tip group (allP 〉 0.05). Either after puncture or after the electrodes were taken out, the time of bleeding [after puncture (minutes): 4.4±2.3 vs. 4.5±3.1, after the electrodes taken out (minutes): 4.1±2.2 vs. 4.8±2.5] and the volume of bleeding [after puncture (mL): 2.8±2.4 vs. 3.2±3.5, after the electrodes taken out (mL): 3.3±1.7 vs. 3.5±2.6] were not significantly different between the two groups (allP 〉 0.05).Conclusions In living dogs with normal heart beat, the manipulation and
出处 《中国中西医结合急救杂志》 CAS 北大核心 2015年第6期627-630,共4页 Chinese Journal of Integrated Traditional and Western Medicine in Intensive and Critical Care
基金 广东省科技计划项目(2008B030301198)
关键词 心内膜起搏 钢丝电极 正常心搏 Endocardiac pacing Steel wire electrode Normal heart beat
  • 相关文献

参考文献11

二级参考文献79

共引文献116

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部