骨水泥聚乙烯内衬固定技术的生物力学研究 被引量:2

Biomechanics study of polyethylene liner cementing into metal acetabular shell
下载PDF
导出
摘要 背景:在因骨溶解和内衬失败进行翻修手术时,用骨水泥将新聚乙烯内衬固定于稳定金属髋臼内是可选择的手术方式。如果金属髋臼外翻或前倾角度不良,单纯更换内衬可有脱位的风险。目的:了解聚乙烯内衬在一定范围内改变接触面积以骨水泥固定于金属髋臼内的早期力学强度。方法:选用40个60 mm外径的金属髋臼、40个28 mm内径的聚乙烯内衬,其中8个聚乙烯内衬外径51 mm为标准锁定组即对照组,32个聚乙烯内衬外径47 mm按内衬与髋臼赤道面交角不同平均分为0°、10°、20°、30°四个实验组。内衬外表面纹理化后以骨水泥固定于金属髋臼内,各组分别进行杠杆及扭转力学实验,观察内衬或内衬-骨水泥界面失败的情况,测量标准锁定装置及骨水泥固定内衬的最大抗杠杆力及抗扭力扭矩。结果:对照组所有内衬均出现分离或移位,实验组各组均未出现内衬-骨水泥界面的失败。4个实验组的最大抗杠杆力及抗扭力扭矩均优于对照组(P<0.05)。实验组各组间最大抗杠杆力及最大抗扭力扭矩的差异无统计学意义(P>0.05)。结论:聚乙烯内衬以骨水泥固定于金属髋臼内可以获得足够的初始固定力学强度。聚乙烯内衬外表面纹理化后,内衬与骨水泥、金属髋臼接触面积减少在一定范围内不会影响内衬-骨水泥界面的初始固定力学强度。 Background: Cementing polyethylene liner into a well-fixed and well-positioned acetabular metal shell has been considered as an alternative procedure in cases of osteolysis or liner failure. If a stable cup was inappropriately inclined or anteverted,liner exchange would be in risk of dislocation.Objective: To evaluate the strength of cementing polyethylene liner into a metal acetabular shell with different contact areas in early term of the procedure.Methods: According to liner fixed methods and different degree versions with respect to the shell, 40 pairs of metal acetabular cups and polyethylene liners were divided into groups. 51 mm outer diameter(OD) liners were fixed in 60 mm ODmetal shells with standard locking mechanism and 47 mm OD liners were cemented into identical shells. Lever- out and torsion tests were performed to observe constructs' fail and measure the early strength of standard locking mechanism and liner-cement interface.Results: All liners with standard locking mechanism failed in control group, while none of liners in experimental groups were disengaged from the cement mantle. Lever- out and torque strength in experimental groups were significantly greater than that in the control group(P〈0.05), but no significant difference was found between four cemented liner groups(P〈0.05).Conclusions: Cementing a polyethylene liner into a well-fixed shell is strong and safe in early fixation. With textured liner,declined contact area between polyethylene and cement will not weaken the liner-cement interface.
出处 《中国骨与关节外科》 2015年第4期348-351,343,共5页 Chinese Journal of Bone and Joint Surgery
基金 广东省科技厅项目(项目编号:2013B021800306) 广州市属高校科技计划项目(项目编号:1201420096) 广州医科大学科研项目(项目编号:2013C36)
关键词 髋关节 翻修 骨水泥 聚乙烯 生物力学 Hip Revision Cement Polyethylene Biomechanics
  • 相关文献

参考文献11

  • 1Haft GF,Heiner AD,Callaghan JJ,et al.Polyethylene liner cementation into fixed acetabular shells.J Arthroplasty,2002,17(4 Suppl 1):167-170. 被引量:1
  • 2Hofmann AA,Prince EJ,Drake FT,et al.Cementation of a polyethylene liner into a metal acetabular shell.J Arthroplasty,2009,24(5):775-782. 被引量:1
  • 3Mauerhan DR,Peindl RD,Coley ER,et al.Cementation of polyethylene liners into well-fixed metal shells at the time of revision total hip arthroplasty.J Arthroplasty,2008,23(6):874-878. 被引量:1
  • 4Beaule PE,Ebramzadeh E,Le Duff M,et al.Cementing a liner into a stable cementless acetabular shell:the doublesocket technique.J Bone Joint Surg Am,2004,86(5):929-934. 被引量:1
  • 5Jiranek WA.Acetabular liner fixation by cement.Clin Orthop Relat Res,2003,417:217-223. 被引量:1
  • 6Springer BD,Hanssen AD,Lewallen DG,et al.Cementation of an acetabular liner into a well-fixed acetabular shell during revision total hip arthroplasty.J Arthroplasty,2003,18(7 Suppl 1):126-130. 被引量:1
  • 7刘志宏,冯建民,杨庆铭.单纯聚乙烯内衬更换在髋关节翻修手术中的应用[J].中华关节外科杂志(电子版),2009,3(5):2-4. 被引量:4
  • 8Koh KH,Moon YW,Lim SJ,et al.Complete acetabular cup revision versus isolated liner exchange for polyethylene wear and osteolysis without loosening in cementless to-tal hip arthroplasty.Arch Orthop Trauma Surg,2011,131(11):1591-1600. 被引量:1
  • 9Blom AW,Astle L,Loveridge J,et al.Revision of an acetabular liner has a high risk of dislocation.J Bone Joint Surg Br,2005,87(12):1636-1638. 被引量:1
  • 10Lombardi AV Jr,Berend KR.Isolated acetabular liner exchange.J Am Acad Orthop Surg,2008,16(5):243-248. 被引量:1

二级参考文献16

  • 1Joseph M,Susan C.Improving the accuracy of acetabular componentorientation:avoiding malposition[J].J Am Acad Orthop Surg,2010,18(5):286-296. 被引量:1
  • 2Dai X,Omori H,Okumura Y,et al.Serial measurement of polyethylenewear of well-fixed cementless metal-backed acetabular component intotal hip arthroplasty:an over 10 year follow-up study[J].Artif organs,2000,24(9):746-751. 被引量:1
  • 3Hayakawa K,Minoda Y,Aihara M.Acetabular component orientation inintra-and postoperative positions in total hip arthroplasty[J].ArchOrthop Trauma Surg,2009,129(9):1151–1156. 被引量:1
  • 4Biedermann R,Tonin A,Krismer M,et al.Reducing the risk ofdislocation after total hip arthroplasty:the effect of orientation of theacetabular component[J].J Bone Joint Surg Br,2005,87(6):762–769. 被引量:1
  • 5Sarmiento A,Latta LL.A radiographic review of 135 total hip Charnleyarthroplasties followed between 15 and 35 years[J].Acta Chir OrthopTraumatol Cech,2006,73(3):145–150. 被引量:1
  • 6Lima DD,Urquhart AG,Burhler KO,et al.The effect of the orientationof the acetabular and femoral components on the range of motion of thehip at different head-neck rations[J].J Bone Joint Surg Am,2000,82(3):315-321. 被引量:1
  • 7Lewinnek GE,Lewis JL,Tarr R,et al.Dislocations after totalhip-replacement arthroplasties[J].J Bone Joint Surg Am,1978,60(2):217-220. 被引量:1
  • 8Widmer KH,Zurfluh B.Compliant positioning of total hip componentsfor optimal range of motion[J].J Orthop Res,2004,22(4):815-821. 被引量:1
  • 9Padgett DE,Hendrix SL,Mologne TS,et al.Effectiveness of anacetabular positioning device im primary total hip arthroplasty[J].HSSJournal.2005:1(1):64-67. 被引量:1
  • 10Moskal JT,Capps SG.Acetabular component positioning in total hiparthroplasty:an evidence-based analysis[J].J Arthroplasty,2011,26(8):1432-1437. 被引量:1

共引文献8

同被引文献13

引证文献2

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部