期刊文献+

冲击电压标准波源的研制 被引量:9

Study on Impulse Voltage Calibrator
下载PDF
导出
摘要 IEC标准规定使用高精度的冲击电压标准波源校准数字记录仪,为了能够校准冲击电阻分压器和数字记录仪组成的完整测量系统,文中介绍了可产生IEC 60060-1—2001中规定的雷电全波的冲击电压标准波源的原理、设计和结构。冲击电压标准波源为小型的MAX发生器,其电压峰值Up、波前时间T1、波尾时间T2都可以根据脉冲形成回路的元器件参数计算得到。该标准波源的输出电压波形峰值为80~1 000 V,理论计算时间参数(0.791/58.22)μs。采用PTB校准过的14Bit数字记录仪测量低阻抗标准的输出波形,电压峰值Up实际测量结果与设置值的偏差在0~0.8‰之间;波前时间T1的偏差-2‰--3‰之间;半峰值时间T2的偏差为-1‰~0.8‰之间,峰值参数满足±1‰的允许误差要求,时间参数均满足±1%的最大允许误差要求,连续重复20次,最大标准偏差为2.5×10-4,一年内各电压点的峰值长期稳定性在0.3‰内。该标准波源可作为冲击电压溯源过程中校准数字记录仪的主标准,校准电阻值为3.25 kΩ的冲击电阻分压器,校准分压比为93.42,结果与分压器频率响应试验结果一致,时间参数与MATLAB程序计算结果吻合,表明该低阻抗冲击电压标准波源也可校准电阻分压器。 According to the requirements of IEC standard, the high accuracy impulse voltage calibrator is used to calibrating the digital recorder. In order to calibrate the complete measuring system composed of impulse resistance divider and digital recorder, the principle, design and construction of a pulse voltage calibrator for generating the lightning impulse according to IEC 60060-1--2001 is introduced. The calibrator is a small MAX generator. And the peak value of voltage Up, front time T1, and time to half value T2 of the generated impulse are calculated from the calibrated values of the components in the impulse-forming circuit. The peak voltage output of the calibrator is from 80 V to 1 000 V, and the theoretical time parameter is (0.791/58.22)μs. The deviations of the calibrated digital recorder are from 0 to 0.8%o for the peak value of voltage, from -2%0 to 3%0 for the front time T1, and from -1‰to 0.8‰ for the time to half value T2, satisfying the accuracy requirements of±1% for the peak value of impulse voltage and less than ±1% for the time parameters. Repeating 20 measurements, the max standard deviation is 2.5×10-4. In whole year, the deviation of the peak value of voltage keeps less than 0.3‰. The calibrator can be used as the primary reference of impulse voltages for calibration of high voltage. An impulse resistance divider with a resistance of 3.25 kΩ is calibrated by the calibrator, and a dividing ratio of 93.42 is obtained, which agrees with the result of frequency response test, and the time parameters also coincide with the theoretical calculation value via MATLAB. It is concluded that the low impedance calibrator can also be applied to calibration of impulse resistance divider.
出处 《高压电器》 CAS CSCD 北大核心 2015年第9期140-145,共6页 High Voltage Apparatus
关键词 冲击电压 溯源 不确定度 稳定性 带载能力 impulse voltage traceability uncertainty stability load capacity
  • 相关文献

参考文献19

  • 1IEC 61083--2001 Instruments and software used for measurement in high-voltage impulse tests[S]. 2001. 被引量:1
  • 2IEEE Standard 1241--2010 IEEE standard for terminolo- gy and test methods for analog-to-digital converters [S]. 2001. 被引量:1
  • 3IEEE Standard 1122TM--1987 IEEE standard for digital recorders for measurements in high-voltage impulse tests [S]. 2001. 被引量:1
  • 4IEEE Standard 1122TM--1998(R2007) IEEE standard for digital recorders for measurements in high-voltage impulse tests[S]. 2001. 被引量:1
  • 5IEEE Standard 4--1995 IEEE standard techniques for high-voltage testing[S]. 2001. 被引量:1
  • 6IEEE Standard 1057--2007 IEEE standard for digitizing waveform recorders[S]. 2001. 被引量:1
  • 7IEC 61083 Ed2.0 42/271/CD software used for measure- ment in high-voltage impulse tests[S]. 2001. 被引量:1
  • 8GB八116896.2-2010高电压冲击测量仪器和软件第2部分:软件的要求[S].2010. 被引量:1
  • 9DL/T992--2006冲击电压测量实施细则[M].北京:中国电力出版社.2006. 被引量:1
  • 10HJi.LLSTROM J K, CHEKUROV Y Y, MARTYI M. A cal- culable impulse voltage calibrator[D]. Helsinki:Helsinki University of technology, 2002. 被引量:1

二级参考文献10

同被引文献103

引证文献9

二级引证文献22

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部