Sysmex XE-5000全自动血液细胞分析仪有核红细胞检测的临床应用评价 被引量:9

Clinical application and evaluation of Sysmex XE-5000 automated blood cell analyser for detecting nucleated red blood cells
下载PDF
导出
摘要 目的评价XE-5000全自动血液细胞分析仪检测外周血有核红细胞(NRBC)的主要性能特点,并探讨其临床应用价值。方法对XE-5000全自动血液细胞分析仪的主要性能进行检测,并选取137份经分析仪DIFF通道检测提示NRBC阳性的患者血液标本,采用分析仪NRBC通道检测NRBC计数和百分比,采用显微镜分类检测NRBC百分比,采用SPSS18.0统计软件进行统计分析。结果高、中和低值标本的NRBC计数及NRBC百分比的批内精密度分别为2.10%、3.26%、11.62%和3.79%、5.80%、13.33%,携带污染率NRBC计数为0.51%,NRBC百分比为0.26%,在(0~18)×109/L范围内XE-5000全自动血液细胞分析仪对NRBC的检测结果呈良好的线性:Y=1.048 6 X+0.189 6(r=0.999 1)。XE-5000全自动血液细胞分析仪与显微镜分类计数法的NRBC百分比检测结果比较差异无统计学意义(P=0.716),呈良好的相关性:Y=1.150 2 X+0.626 1(r=0.967 0)。结论 XE-5000全自动血液细胞分析仪可替代传统的显微镜分类计数方法应用于临床。 Objective To evaluate the performance of XE‐5000 automated blood cell analyser for detecting nucleated red blood cell (NRBC) in peripheral blood and investigate its clinical application value .Methods The intra‐assay imprecision ,carryover rate , and linear range of the analyser were evaluated .The absolute NRBC count and percentage of NRBC of 137 blood specimens (NRBC‐positive according to the DIFF channel of the analyser) were determined in the NRBC channel of the analyser ,and the percentage of NRBC of these blood specimens were determined by using microscope method as well .Differences between the two methods were analysed by using SPSS18 .0 statistic software .Results The intra‐assay imprecision of the analyser for detecting absolute NRBC count in specimens with high ,moderate ,and low Q‐flag values were 2 .10% ,3 .26% and 11 .62% ,respectively ,and the imprecision for detecting percentage of NRBC were 3 .79% ,5 .80% and 13 .33% ,respectively .The carryover rates of the analyser for detecting absolute NRBC count and percentage of NRBC were 0 .51% and 0 .26% ,respectively .At the range of (0~18)×109/L ,the absolute NRBC count detected by using the analyser showed good linearity :Y=1 .048 6X+0 .189 6(r=0 .999 1) .There were no statistically significant differences between the analyser and microscope for detecting percentage of NRBC(P=0 .716) ,and showed good corre‐lation:Y=1 .150 2X+0 .626 1(r=0 .967 0) .Conclusion The XE‐5000 automated blood cell analyser could completely replace the traditional microscope for clinically classifying and counting NRBCs .
出处 《国际检验医学杂志》 CAS 2015年第18期2666-2668,2671,共4页 International Journal of Laboratory Medicine
关键词 血液细胞分析仪 有核红细胞 性能 外周血 blood cell analyzer nucleated red blood cell performance peripheral blood
  • 相关文献

参考文献22

  • 1de Keijzer MH,van der Meer W. Automated counting of nucleated red blood cells in blood samples of newborns[J]. Clin Lab Haema tol, 2002,24 (6) : 343-345. 被引量:1
  • 2Ghosh B, Mittal S,Kumar S, et al. Prediction of perinatal asphyxia with nucleated red blood cells in cord blood of newborns[J]. Int J Gynaecol Obstet,2003,81(3) :267-271. 被引量:1
  • 3Hanlon-l.undberg KM, Kirby RS. Nucleated red blood cells as a marker of aeldemia in term neonates[J]. Am J Ohstet Gyneeol, 1999,181(1) :196-201. 被引量:1
  • 4Stachon A, Kempf R, Holland-Letz T, et al. Daily monitoring of nucleated red blood cells in the blood of surgical intensive care pa- tients[J]. Clin Chim Acta,2006,366(1/2) :329-335. 被引量:1
  • 5Stachon A, Sondermann N, Imohl M, et al. Nucleated red blood cells indicate high risk of in-hospital mortality[J]. J I.ab Clin Med,2002,140(6) :407-412. 被引量:1
  • 6Urrechaga E,Borque L,Escanero JF. Potential utility of the new Sysmex XE 5000 red blood cell extended parameters in the study of disorders of iron metabolism[J]. Clin Chem Lab Med, 2009,47 (11) :1411-1416. 被引量:1
  • 7Stachon A, Segbers E, H olland-Letz T, et al. Nucleated red blood cells in the blood of medical intensive care patients indicate in creased mortality risk: a prospective cohort study[J]. Crit Care, 2007,11 (3) :R62. 被引量:1
  • 8Schoorl M,Gaag DVD,Schoorl M,et al. Changes in red blood cell hemoglobinization during pregnancy[J]. Ned Tijdschr Klin Chem I.abgeneesk, 2010, 35 : 206-208. 被引量:1
  • 9Clinical and Laboratory Standards Institute. H20-A2 Reference leukocyte(WBC) differential count (proportional) and evaluation of instrumental methods[S]. 2nd cd. Wayne, PA : CLSI, 2008. 被引量:1
  • 10Buttarello M. Quality specification in hacmatology:the automated blood cell count[J]. Clin Chim Acta, 2004,346 ( 1 ) : 45-54. 被引量:1

同被引文献61

引证文献9

二级引证文献33

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部