不同形状超声乳化针头在扭动超声模式的应用 被引量:2

Application of different shape tips in torsional microcoaxial phacoemulsification
下载PDF
导出
摘要 目的:通过与Infinity超声乳化针头对比,观察Centurion双重弯曲形状设计的超声乳化针头的临床效果。方法:前瞻性对照研究,将年龄相关性白内障患者132例132眼,随机分为两组,试验组66例66眼采用Centurion超声乳化针头,对照组66例66眼采用Infinity超声乳化针头,两组均用Centurion超声乳化仪在同一手柄下,利用相同参数下完成常规的超声乳化联合人工晶状体植入术。记录两组术中的累积消耗能量(cumulative dissipated energy,CDE),观察术后1wk,1mo的最佳矫正视力(best corrected visual acuity,BCVA)、角膜内皮密度(corneal endothelial cell density,CD)、中央角膜厚度(central corneal thickness,CCT)。采用独立样本t检验和χ2检验进行统计分析。结果:两组患者的CDE差异有统计学意义(P<0.05)。术后1wk时BCVA(Log MAR)试验组优于对照组,差异有统计学意义(P<0.05)。术后1wk试验组CCT小于对照组,差异有统计学意义(P<0.01),但术后1mo两组无统计学差异(P>0.05)。结论:Centurion超声乳化针头能有效减少术中超声能量的释放,减少术后角膜内皮的损伤,对术后早期视力恢复更具有优势。 AIM:To compare the clinical outcomes of torsional microcoaxial phacoemulsification by using different shape tips,including Centurion ultrasound tips and Infinity tips.METHODS:A randomized prospective study was conducted.132 patients(132 eyes) with age- related cataract,were randomly divided into study group(Centurion ultrasound tips) and control group(Infinity ultrasound tips),66 cases(66 eyes) in each group.Under the same parameter of the Centurion phacoemulsification instrument,all the patients received normal microcoaxial phacoemulsification and intraocular lens implantation.The cumulative dissipated energy(CDE) was recorded during the operation.The best corrected visual acuity(BCVA),corneal endothelial cell density(CD) and central corneal thickness(CCT) were compared at 1wk and 1mo after surgery.Statistical analysis of the data was performed using Student's t test and Chi-square test.RESULTS:There was significant difference on CDE(P〈0.05) between these two groups.On 1wk after surgery,the study group showed better BCVA(LogMAR) as compared that of control group,the difference was statistically significant(P〈0.05).The CCT of study group was smaller than the control group on 1wk after surgery(P〈0.01),but there was no significant difference on 1mo(P〉0.05).CONCLUSION:Centurion ultrasound tips can significantly reduce the cumulative dissipated energy and cause less damage to corneal endothelial cells.These findings suggest that centurion vision system phacoem ulsification surgery could get earlier visual rehabilitation postoperative.
出处 《国际眼科杂志》 CAS 2015年第8期1441-1443,共3页 International Eye Science
关键词 超声乳化白内障吸除术 白内障 累计消耗能量 超声乳化针头 phacoemulsification cataract cumulative dissipated energy ultrasound tips
  • 相关文献

参考文献9

  • 1Sharif-Kashani P, Fanney D, Injev V. Comparison of occlusion break responses and vacuum rise times of phacoemulsification systems, BMC Ophthalmol 2014 , 14:96. 被引量:1
  • 2Christakis PG, Braga- Mele RM. Intraoperative performance and postoperative outcome comparison of longitudinal, torsional and transversal phacoemulsification machines. J Cataract Refract Surg 2012, 38(2) :234-241. 被引量:1
  • 3Kim DH, Wee WR, Lee JH, et al . The comparison between torsional and conventional mode phacoemulsification in moderate and hard cataracts. Korean J Ophthalmol 2010 ,24 ( 6 ) :336-340. 被引量:1
  • 4Farukhi AM, Stagg BC, Ronquillo C Jr, et al. Effect of phaco tip diameter on efficiency and chatter. J Cataract Refract Surg 2014,40(5) : 811-817. 被引量:1
  • 5Helvacioglu F, Sencan S, Yeter C, et al. Outcomes of torsional microcoaxial phacoemulsification using tips with 30- degree and 45- degree aperture angles. J Cataract Refract Surg 2014 ,40 ( 3 ) :362-368. 被引量:1
  • 6McNeill JI. Flared phacoemulsification tips to decrease ultrasound time and energy in cataract surgery. J Cataract Refract Surg 2001 , 27 ( 9 ) : 1433-1436. 被引量:1
  • 7Mencucci R, Ponchietti C, Virgili G, et al. Corneal endothelial damage after cataract surgery: mieroincision versus standard technique. J Cataract Refract Surg 2006 ,32 ( 8 ) : 1351 - 1354. 被引量:1
  • 8Mahdy MA, Eid MZ, Mohammed MA, et al. Relationship between endothelialcell loss and microcoaxial phacoemulsification parameters in noneomplieated cataract surgery. Clin Ophthalmol 2012 ,6:503-510. 被引量:1
  • 9姚克,王玮,吴炜,汤霞靖,李兆春,金冲飞.同轴1.8mm微切口超声乳化白内障手术临床效果评价[J].中华眼科杂志,2011,47(10):903-907. 被引量:76

二级参考文献19

  • 1Yao K, Tang X J, Ye PP. Corneal astigmatism, high order aberrations, and optical quality after cataract surgery : microincision versus small incision. J Cataract Refract Surg,2006, 22 : S1079-1082. 被引量:1
  • 2Can I, Takmaz T, Yildiz Y, et al. Coaxial, microcoaxial, and 5iaxial microincision cataract surgery: prospective comparative study. J Cataract Refract Surg, 2010,36 : 740-746. 被引量:1
  • 3Berdahl JP, DeStafeno JJ, Kim T. Corneal wound architecture and integrity after phacoemulsification evaluation of coaxial, microincision coaxial, and microincision bimanual techniques. J Cataract Refract Surg,2007,33: 510-515. 被引量:1
  • 4Emery JM, Little JH. Phacoemulsification and aspiration of cataracts ; Surgical techniques, complications, and results. St Louis, MO, CV Mosby, 1979:45-48. 被引量:1
  • 5Alio JL, Rodriguez-Prats JL, Vianello A, et al. Visual outcome of microincision cataract surgery with implantation of an Acfi. Smart lens. J Cataract Refract Surg, 2005,31 : 1549-1556. 被引量:1
  • 6Vasavada V, Raj SM, Vasavada AR. Intraoperative performance and postoperative outcomes of microcoaxial phacoemulsification. Observational study. J Cataract Refract Surg,2007,33: 1019-1024. 被引量:1
  • 7Wilczynski M, Supady E, Piotr L, et al. Comparison of surgically induced astigmatism after coaxial phacoemulsification through 1.8 mm microincision and bimanual phacoemulsification through 1.7 mm microincision. J Cataract Refract Surg, 2009,35: 1563-1569. 被引量:1
  • 8Wilczynski M, Supady E, Loba P, et al. Comparison of early corneal endothelial cell loss after coaxial phacoemulsification through 1.8 mm microincision and bimanual phacoemulsification through 1.7 mm microincision. J Cataract Refract Surg, 2009,35 : 1570-1574. 被引量:1
  • 9Yentura AC, Walti R, Bohnke M. Corneal thickness and endothelial density before and after cataract surgery. Br J Ophthalmol,200l, 85 : 18-20. 被引量:1
  • 10Walkow T, Anders N, Klebe S. Endothelial cell loss after phacoemulsification: relation to preoperative and intraoperative parameters. J Cataract Refract Surg ,2000,26 : 727-732. 被引量:1

共引文献75

同被引文献13

引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部