摘要
目的:评价Scoligauge作为脊柱侧凸筛查工具的可靠性。方法:6名具有多年脊柱侧凸筛查经验的医师和治疗师分别利用安装Scoligauge软件的i Phone手机和传统的脊柱测量仪脊柱侧凸测量仪(scoliometer)对27例确诊为特发性脊柱侧凸的青少年患者,男10例,女17例,平均年龄12.8岁,进行躯干倾斜角度的测量,每例患者接受每周1次,共4次测量。通过分析两种测量手段的观测值差异和95%置信区间,以及组内相关系数来评估Scoligauge的测量信度。结果:Scoligauge与脊柱侧凸测量仪测量得到的躯干倾斜角配对绝对差为2.5°±1.4°,差异的95%置信区间为±2.9°。Scoligauge的组内相关系数值为0.915,脊柱侧凸测量仪的组内相关系数值为0.932。结论:Scoligauge与脊柱侧凸测量仪在脊柱侧凸测量上可视为等效工具。
Objective: To evaluate the reliability of rib hump measurement in scoliosis with Scoligauge.
Method: Six experienced doctors and physiotherapists (observer) measured the rib hump angles of the 27 pa- tients (10 males and 17 females with average age of 12.8yr.) with adolescent idiopathic scoliosis using the sco- liometer and the iPhone with Scoligauge application software. Every patients received one measurement per week and total four measurements performed in four weeks. Measurement difference between pairs of iPhone/ scoliorneter measurements and 95% confidence intervals were calculated. The intraclass correlation coefficients (ICC) were calculated for repeated measurements of each patient by observers.
Result: Absolute difference of rib hump measurements between pairs of Scoligauge/scoliometer was (2.5±1.4) degrees, the 95% confidence interval for differences on the same subject was 4-2.9 degrees. The ICC were 0.915 (Scoligauge) and 0.932 (scoliometer).
Conclusion: The iPhone using Scoligauge is a clinically equivalent rib hump measurement tool to the scoliome- ter in spinal deformity patients.
出处
《中国康复医学杂志》
CAS
CSCD
北大核心
2015年第3期247-250,共4页
Chinese Journal of Rehabilitation Medicine
基金
上海市科学技术委员会资助项目(13401903400)
崇明县科学技术委员会资助项目(CKY2013-03)
新华人才基金项目(12XJ22001)