小直径种植体折断原因分析及文献复习 被引量:2

Analysis of reduced diameter implant fracture and literature review
下载PDF
导出
摘要 近年来对小直径种植体的临床应用和系统研究逐渐增多。该文报告上颌3颗ITI 3.3 mm种植体折断病例1例,结合临床分析及相关文献可认为在特殊条件下如咬合负担过重,种植体存在机械并发症(如金属疲劳,缝隙腐蚀及氢脆)时,小直径种植体机械强度方面的不足可能更容易导致种植体折断的发生。 The application and studies of reduced diameter implants(RDI)are increasing recently.A case with fracture of three 3.3 mm ITI implants in maxilla was reported in this paper.Combined with the related literature,overload,mechanical complications(such as fatigue, crevice corrosion and hydrogen embrittlement)and the insufficiency of mechanical properties of RDI may lead to the fracture.
出处 《实用口腔医学杂志》 CAS CSCD 北大核心 2015年第1期133-136,共4页 Journal of Practical Stomatology
基金 国家自然科学基金(编号:81170998) 广东省医学科研基金(编号:B2012032)
关键词 小直径种植体 折断 应力 疲劳 缝隙腐蚀 氢脆 Reduced diameter implant Fracture Stress Fatigue Crevice corrosion Hydrogen embrittlement
  • 相关文献

参考文献6

二级参考文献41

  • 1巢永烈,杨永丰,赵云凤.天然牙-末端种植牙固定桥的受载分析[J].华西口腔医学杂志,1995,13(1):6-10. 被引量:55
  • 2孔亮,刘宝林,胡开进,李德华,宋应亮,马攀,杨劲.螺纹种植体螺距的优化设计和应力分析[J].华西口腔医学杂志,2006,24(6):509-512. 被引量:23
  • 3马攀,刘洪臣,李德华,林升,彭勤建,史真,吴晓霞.三种螺距对种植体初期稳定性影响的有限元研究[J].口腔颌面修复学杂志,2007,8(1):47-49. 被引量:20
  • 4王惠芸.我国人牙的测量和统计[J].中华口腔科杂志,1959,7(3):149-149. 被引量:173
  • 5-.齿科技工[M].医齿药出版株式会社,1991(3).283. 被引量:1
  • 6Lang NP, Berglundh T, Heitz-Mayfield LJ, et al. Consensus statements and recommended clinical procedures regarding implant survival and complications[ J]. Int J Oral Maxillofac hnplants, 2004, 19(suppl) : 150 - 154. 被引量:1
  • 7van Zyl PP, Grundling NL, Jooste CH, et al. Three-dimensional finite element model of a human mandible incorporating six osseointegrated implants for stress analysis of mandibular cantilever prostheses[ J ]. Int J Oral Maxillofac Implants, 1995, 10(1) : 51 -57. 被引量:1
  • 8Van Rossen IP, Braak LH, de Putter C, et al. Stress-absorbing elements in dental implants. [ J]. J Prosthet Dent, 1990, 64(2) : 198 -205. 被引量:1
  • 9Hou SM, Hsu CC, Wang JL, et al. Mechanical tests and finite element models for bone holding power of tibial locking screws[J]. Clin Biomech (Bristol, Avon), 2004, 19(7): 738 - 745. 被引量:1
  • 10Tepper G, Haas R, Zechner W, et al. Three-dimensional finite element analysis of implant stability in the atrophic posterior maxilla: A mathematical study of the sinus floor augmentation[ J]. Clin Oral Implants Res, 2002, 13 (6) : 657 - 665. 被引量:1