免疫比浊法检测梅毒螺旋体抗体的方法评价 被引量:2

Method Evaluation on Serum Concentration of Anti-treponema Pallidum Antibody by Immunoturbidimetry
下载PDF
导出
摘要 目的评价免疫比浊法在梅毒螺旋体抗体检测中的临床应用价值。方法分析免疫比浊法的精密度、线性、稳定性、干扰因素、携带污染指标,并与化学发光微粒子免疫分析法(CMIA)进行阳性率比较。结果免疫比浊法批内CV为1.64%~2.22%、批间CV为3.57%~3.74%;黄疸、脂血及溶血对本方法无明显干扰;阳性率均为16.8%,差异无统计学意义(x^2=0,P>0.05)。结论免疫比浊法检测结果与CMIA法检测结果相关性良好,可以满足临床诊断需要。 Objective To evaluate clinical value of TP-Ab measurement by immunoturbidimetry. Methods The precision, linearity, stability, interfere factor and carryover in TP-Ab detection were analyzed. The positive rate of TP-Ab detection with immunoturbidimetry compared with CMIA was also analyzed. Results The within-run CV of immunoturbidimetry was 1.64%-2.22%. The between-run CV was 3.57%-3.74%. Icterus, lipemia and hematolysis had no significant interference on immunoturbidimetry measurement. The positive rate with both method were 16.8%, (X2= 0 ,P 〉 0.05 ). Conclusion The results of immunoturbidimetry are good correlation compared with CMIA, which is suitable for clinical application.
出处 《标记免疫分析与临床》 CAS 2015年第1期56-58,共3页 Labeled Immunoassays and Clinical Medicine
基金 解放军总医院临床科研扶持基金(NO.2012FC-TSYS-3054)
关键词 免疫比浊法 化学发光微粒子免疫分析法 梅毒螺旋体抗体 脂血 溶血 Immunoturbidimetry CMIA TP- Ab Lipemia Hemolytic