摘要 腔内隔绝术用人工血管(stent-graft,SG)长期植入人体后会出现不同程度疲劳损坏而导致内漏.针对这一问题设计制备了两种不同机织覆膜组织结构的SG试样,采用体外扭转弯折疲劳模拟装置对其进行体外仿真疲劳模拟,并对疲劳后试样进行非破坏性(总体观察)和破坏性(织物结构、物理特性)测试研究.结果表明:随着实验时间的延长,SG疲劳现象加剧;覆膜、支架及缝合线都出现了不同程度的老化破损;覆膜结晶度下降;覆膜组织结构为平纹的SG比3/1斜纹的SG更容易产生疲劳破损,但其结构更加稳定. Long-term complications after endovascular aortic repair(EVAR)including endoleaks arise from a defect in the stent-graft(SG).In order to solve this problem,two different weaving structures of SG were designed and tested in a long-term in vitro buckling and twisting fatigue simulation environment to investigate the fatigue performance.Also,the non-destructive(gross observations)and destructive(fabric characteristics,mechanical properties)properties were analyzed.The results showed that,as the testing time prolonged,the fatigue performance of SG was aggravated.There were different degrees of aging damages in the fabrics,stents and sutures.Also the crystallinity of the fabrics declined.The SG fitted with plain weave structure showed a considerably lower endurance but a higher structural stability compared to those fitted with 3/1 twill weave structure.
出处 《东华大学学报(自然科学版)》 CAS CSCD 北大核心 2014年第6期681-686,共6页 Journal of Donghua University(Natural Science)
基金 上海高校实验技术队伍建设计划资助项目(101-07-0053014) 中央高校基本科研业务费专项资金资助项目(14D110132)
关键词 腔内隔绝术用人工血管 织物组织 体外疲劳模拟 疲劳性能 stent-graft weave structure in vitro fatigue simulation fatigue performance
  • 相关文献

参考文献3

二级参考文献4

  • 1Alexandra Rodrigues,Lígia Figueiredo,Jo?o Bordado.Abrasion behaviour of polymeric textiles for endovascular stent-grafts[J].Tribology International.2012 被引量:1
  • 2Nicolas Demanget,Stéphane Avril,Pierre Badel,Laurent Orgéas,Christian Geindreau,Jean-No?l Albertini,Jean-Pierre Favre.Computational comparison of the bending behavior of aortic stent-grafts[J].Journal of the Mechanical Behavior of Biomedical Materials.2011(1) 被引量:1
  • 3Anders Wanhainen,Rickard Nyman,Mats-Ola Eriksson,Martin Bj?rck.First report of a late type III endoleak from fabric tears of a Zenith stent graft[J].Journal of Vascular Surgery.2008(3) 被引量:1
  • 4Tikva S. Jacobs,Jamie Won,Edwin C. Gravereaux,Peter L. Faries,Nicholas Morrissey,Victoria J. Teodorescu,Larry H. Hollier,Michael L. Marin.Mechanical failure of prosthetic human implants: A 10-year experience with aortic stent graft devices[J].Journal of Vascular Surgery.2003(1) 被引量:1

共引文献3

同被引文献12

引证文献2

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部