期刊文献+

第二代双源CT的Flash序列在肋骨外伤检查中的应用价值

Application Value of Flash Sequence of Second-Generation Dual-Source CT in the Examination of Rib Injury
下载PDF
导出
摘要 目的探讨第二代双源CT的Flash序列在肋骨外伤检查中的应用。方法对28例肋骨外伤患者分别使用常规螺旋序列扫描和二代双源Flash序列扫描,比较两种扫描方法的图像质量及辐射剂量。结果常规螺旋序列扫描图像的优良率明显低于Flash序列扫描图像。Flash序列扫描的管电流、CT容积剂量指数、剂量长度乘积、有效剂量、扫描时间均低于常规螺旋扫描,差异具有统计学意义(P<0.05)。结论 Flash序列扫描具有辐射剂量低、扫描时间短、图像质量高等特点,优于常规螺旋序列扫描,值得推广。 Objective To explore the application value of Flash sequence of second-generation dual-source CT in the examination of rib injury. Methods 28 patients with rib injury were scanned by both conventional spiral sequence scan and second-generation dual-source Flash sequence scan. Then the image quality and radiation dose of the two methods were compared. Results The high-quality image rate of conventional spiral sequence scan were obviously lower than that of Flash sequence scan;the tube current, volume CT dose index (CTDIvol), dose-length product (DLP), effective dose (ED) and scan time of Flash sequence scan were signiifcantly lower than those in conventional spiral sequence scan (P〈0.05). Conclusion Flash sequence scan is better than conventional spiral sequence scan in radiation doses, scanning time and image quality which is worthy to be widely applied.
出处 《中国医疗设备》 2014年第10期139-141,共3页 China Medical Devices
关键词 双源CT 常规螺旋序列 Flash序列 肋骨外伤 dual-source CT conventional spiral sequence Flash sequence rib injury
  • 相关文献

参考文献8

二级参考文献30

  • 1孙丛,王锡明,徐卓东,邓凯,柳澄,王道平,段艳华.双能CT在手创伤性疾病中的诊断价值[J].中华临床医师杂志(电子版),2011,5(7):1980-1982. 被引量:3
  • 2储成凤,杨明,刘斌,徐秋贞,杨小庆,朱晓莉,朱亚彬,姚青,陈峰,张俭.动态CT增强扫描对肺结节的定性价值研究[J].临床放射学杂志,2005,24(1):29-32. 被引量:32
  • 3Lell M, Marwan M, Schepis T, et al. Prospectively ECG- triggered high-pitch spiral acquisition for coronary CT angiographyusing dual source CT: techniqueand initial experience. Eur Radiol, 2009,19:2576-2583. 被引量:1
  • 4Achenbach S, Matwan M, Ropers D, et al. Coronary computed tomography angiography with a consistent dose below 1 rosy using prospectively electrocardiogram-triggered high-pitch spiral acquisition. Eur Heart J,2010,31:340-346. 被引量:1
  • 5Aehenbaeh S, GoroU T, Seltmann M, et al. Detection of coronary artery stenoses by low-dose prospectively ECG-triggered high-pitch spiral coronary CT angiography. JACC Cardiovasc Imaging, 2011, 4:328-337. 被引量:1
  • 6Hall EJ. Lessons we have learned from our children:cancer risks from diagnostic radiology. Pediatr Radin1,2002 ,32 :700-706. 被引量:1
  • 7Thomas KE, Wang B. Age-specific effective doses for pediatric MSCT examinations at a large children's hospital using DLP conversion coefficients : a simple estimation method. Pediatr Radiol, 2008,38:645 -656. 被引量:1
  • 8Winklehner A, Karlo C, Puippe G, et al. Raw data-based iterative reconstruction in body CTA: evaluation of radiation dose saving potential. Eur Radiol,2011,21:2521-2526. 被引量:1
  • 9Tsai iC, Lee T, Chen MC, et al. Visualization of neonatal coronary arteries on multidetector row CT: ECG-gated versus non- ECG-gated technique. Pediatr Radiol, 2007, 37: 818-825. 被引量:1
  • 10Achenbach S, Marwan M, Schepis T, et al. High-pitch spiral acquisition: a new scan mode forcoronary CT angiography. J Cardiovasc Comput Tomogr,2009,3 : 117-121. 被引量:1

共引文献31