两种分析仪检测β-HCG的比对结果 被引量:1

Comparison of β-HCG Detection Results by Two Analyzers
下载PDF
导出
摘要 目的探讨两种分析仪测定β-HCG结果的可比性,为本结果的一致性提供依据。方法分别用雅培Architect i1000与西门子Dimension XPand测定β-HCG,依据CLSI的EP9-A2文件规定,以Architect i1000为比较方法,以Dimension XPand为实验方法,进行比对试验。结果两种分析仪结果间均呈良好的线性关系,其相关系数r均大于0.975,相对偏倚小于可接受限(20%)。结论两种分析仪可以用来检测β-HCG,两种分析仪间检验结果有良好的相关性。 Objective To explore the comparability of β--HCG detection results by two analyzers. Methods The serum β--HCG levels were detected by ABBOTT Architect i1000 and SIEMENS Dimension XPand analyzer. The results were compared by referring to the document of NCCLS EPg-A2. Results The testing results of serum β- HCG by two different analyzers were linear correlation (r 〉 0. 975 ). Their anticipated biases were within accepted biases ( 20% ). Conclusion The two analyzers could be used for detection of serum β-- HCG with clinical acceptance.
出处 《标记免疫分析与临床》 CAS 2014年第4期447-449,共3页 Labeled Immunoassays and Clinical Medicine
关键词 Β-HCG DIMENSION XPand ARCHITECT i1000 比对分析 EP9—2A β- HCG Dimension XPand Architect i1000 Comparative Analysis EP9-2A