摘要 目前吸收剂量的测量方法有多种,对低能电子束吸收剂量进行测量时,由于受到电子束穿透深度、热效应、动态加工等因素的影响,可用方法很少,通常用显色薄膜剂量片进行测量.本文采用FWT-60-00薄膜显色剂量片测量1MeV低能电子束进行辐照加工时的吸收剂量,研究了吸收剂量与束流强度、辐照时间、束下装置运行速度的关系,测量了材料中吸收剂量的深度分布,并对加速器能量进行了校正.结果表明:在相同的辐照时间和电子能量下,吸收剂量随束流的增加线性增长;速度较高时,吸收剂量与小车速度也呈现出良好的线性关系. There were many methods to measure the absorbed dosage,while due to the effect of low penetrating depth,thermal effects,dynamic processing,and other factors few methods were available to measure the radiation dosage of electron irradiation with low energy,the color film dosimeter was usually used to measure.In this paper,the radio chromic film dosimeter(FWT-60-00thin films)was used to measure the dose of low energy electron beam at the energy of 1 MeV used for irradiation processing. The relation between the radiation dosage and the beam intensity,irradiation time and the speed of the beam device was investigated,and the depth distribution of irradiation dose was measured,the energy of the accelerator was corrected.The results showed that,with the same time and the electron energy,the absorbed dose was grown linearly with the beam;the dosage and the speed of beam device showed a good linear relationship at high speed.
出处 《测试技术学报》 2014年第3期225-229,共5页 Journal of Test and Measurement Technology
基金 湖北科技学院核技术专项基金资助项目(ZX1103 ZX1022 ZX1102 PY1103 BK1103) 湖北省教育厅科学技术研究资助项目(Q20122808) 国家科技部973计划资助项目(2009CB939705) 国家自然科学基金资助项目(20873096 20921062) 教育部-民委共建湖北省催化材料重点实验室资助项目(CHCL09007)
关键词 薄膜显色剂量片 辐照剂量 束下装置 电子加速器 radiochromic film dosimeter radiation dosage beam device electron accelerator
  • 相关文献

参考文献3

二级参考文献12

  • 1黄扬,杨勇,张圆月,鄢铃,魏川.X、γ射线照射量(治疗水平)标准装置技术研究[J].中国测试技术,2005,31(6):32-35. 被引量:6
  • 2ICRU Report 24:1976,Determination of absorbed dose in a patient irradiated by beams of X or gamma rays in radiotherapy procedures[S].International Commission on Radiation Units and Measurements Bethesda,ICRU,1976. 被引量:1
  • 3IAEA Technical Reports Series No 277:1996,Absorbed dose determination in photon and electron beams:an international code of practice[S].2nd Edition.Vienna,IAEA,1996. 被引量:1
  • 4IAEA Technical Reports Series No 277:1998,Absorbed dose determination in photon and electron beams:an adaptation of the IAEA international codes of practice[S].2nd Edition.Vienna,IAEA,1998. 被引量:1
  • 5IAEA Technical Report Series No 381:1997,The use of plane-parallel ionization chambers in high-energy electron and photon beams.An International Cede of Practice for Dosimetry[S].Vienna,IAEA,1997. 被引量:1
  • 6IAEA Technical Report Series No 398:2000,Absorbed dose determination in external beam radiotherapy.An International Cede of Practice for Dosimetry Based on Standards of Absorbed Dose to Water[S].Vienna,IAEA,2000. 被引量:1
  • 7IAEA Technical Report Series No 398:2004,An International code of preactice for Dosimetry based on absorbed dose to water,Absorbed dose determination in external beam radiotherapy[S].Vienna,IAEA,2004. 被引量:1
  • 8JJG912-1996,治疗水平电离室剂量计[S].北京:中国计量出版社,1997. 被引量:1
  • 9JJG 589-2001,外照射治疗辐射源[S].北京:中国计量出版社,2001. 被引量:1
  • 10.IAEA Techn ical Reports Series No 277 Absobed dose de-term ination in photon and electron beam[].An InternationalCode of IPractice.1997 被引量:1

共引文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部