期刊文献+

应用LabVIEW的聚焦声场测量系统 被引量:2

A Focused Sound Field Measurement System by LabVIEW
下载PDF
导出
摘要 该文根据聚焦声场测量的要求,应用LabVIEW虚拟仪器平台建立了一个聚焦声场测量系统。该系统能够自动寻找聚焦声场的焦点位置,并根据焦域的大小调整扫描路径,三维声场扫描时间从均匀步长的888 h减少到变步长的9.25 h,提高了声场测量效率。理论计算与系统测量结果存在一定偏差,其中焦平面-6 dB宽度差异率为3.691%,声束轴-6 dB长度的差异率为12.937%。 In this paper, according to the requirement of the focused sound field measurement, a focused sound field measurement system was established based on the LabVIEW virtual instrument platform. The system can automatically search the focus position of the sound field, and adjust the scanning path according to the size of the focal region. Three-dimensional sound field scanning time reduced from 888 hours in uniform step to 9.25 hours in variable step. The efficiency of the focused sound field measurement was improved. There is a certain deviation between measurement results and theoretical calculation results. Focal plane - 6 dB width difference rate was 3.691%, the beam axis - 6 dB length differences rate was 12.937%.
出处 《中国医疗器械杂志》 CAS 2014年第3期181-185,共5页 Chinese Journal of Medical Instrumentation
基金 上海市科技攻关项目(11441900200)
关键词 声场测量 系统软件 LABVIEW sound field measurement, system software, LabVIEW
  • 相关文献

参考文献2

二级参考文献15

  • 1冯若 吴至诚.抗癌热疗法的新进展[J].自然杂志,1985,8(9):666-669. 被引量:4
  • 2曹友清 伍烽 等.高强聚焦超声治疗原发性肝癌的病理学研究[J].临床超声医学杂志,2000,2:250-250. 被引量:2
  • 3冯若.从国际声学会议看高强聚焦超声治疗学的兴起[J].中国超声医学杂志,1999,15(3):239-239. 被引量:1
  • 4冯若.我国自主开发的“海扶”(HIFU)聚焦超声刀开始走向世界[J].中国超声医学杂志,2000,16(2). 被引量:1
  • 5冯若,世界医疗器械,1998年,4卷,1期,16页 被引量:1
  • 6冯若,超声诊断设备原理与设计,1993年,906页 被引量:1
  • 7冯若,自然杂志,1985年,8卷,9期,666页 被引量:1
  • 8曹友清,临床超声医学杂志,2000年,2卷,增刊,250页 被引量:1
  • 9冯若,中国超声医学杂志,2000年,16卷,2期,目录后页 被引量:1
  • 10冯若,中国超声医学杂志,1999年,15卷,3期,239页 被引量:1