摘要 目的比较IMSureQA和MatriXX两种计划验证工具,探讨IMSureQA软件在临床剂量验证中的可靠性。方法按随机抽样的抽签法选取10例调强放疗患者,分别用IMSureQA和MatriXX进行剂量验证,比较IMSureQA和MatriXX按3%,3mm标准进行分析后通过率的差异。结果IMSureQA验证方式经剂量计算模型得出,MatriXX的验证通过实际测量获得,验证的治疗计划以3%,3mm标准经y分析后得到其通过率,MatriXX的平均通过率为(97.9±0.6)%,IMSureQA为(98.1±0.8)%(t=0.86,P〉0.05)。结论IMSureQA可用来执行调强计划的验证。 Objective To compare IMSure QA and MatriXX for intensity modulated radiotherapy plans, and to investigate the credibility of IMSure QA software. Methods Ten patients receiving intensity modulated radiotherapy were randomly chosen for the quality assurance plans with 1MSure QA and MatriXX. Results The overall average of percentage pass points in 3% and 3 mm were (98. l ± 0.8)% with IMSure QA, and (97.9 ±0.6)% with MatriXX(t =0.86,P 〉0.05). Conclusion IMSure QA can be a reliable verification tool for IMRT QA.
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2014年第4期292-294,共3页 Chinese Journal of Radiological Medicine and Protection
关键词 调强放射治疗 剂量验证 IMSure QA MatriXX Intensity modulated radiotherapy Dose verification IMSure QA MatriXX
  • 相关文献

参考文献9

  • 1Yang Y, Xing L, Boyer AL, et al. A three-source model for the calculation of head scatter factors [ J ]. Med Phys, 2002, 29 (9) : 2024-2033. 被引量:1
  • 2Yang Y, Xing L, Li JG, et al. Independent dosimetric calculation with inclusion of head scatter and MLC transmission for IMRT[ J]. Med Phys, 2003, 30( 11 ) :2937-2947. 被引量:1
  • 3Low DA, Harms WB, Mutic S, et al. A technique for the quantitative evaluation of dose distributions [ J ]. Med Phys, 1998, 25(5) :656-661. 被引量:1
  • 4Lotourneau D, Gulam M, Yan D, et al.' Evaluation of a 2D diode array for IMRT quality assurance [ J]. Radiother Oncol, 2004, 70(2) :199-206. 被引量:1
  • 5Arjomandy B, Sahoo N, Ciangaru G, et al. Verification of patient-specific dose distributions in proton therapy using acommercial two-dimensional ion chamber array [ J ]. Med Phys, 2010, 37(11) :5831-5837. 被引量:1
  • 6Cheong KH, Kang SK, Lee M, et al. Evaluation of delivered monitor unit accuracy of gated step-and-shoot IMRT using a two- dimensional detector array [ J]. Med Phys, 2010, 37 (3) : 1146-1151. 被引量:1
  • 7Wolfsberger LD, Wagar M, Nitsch P, et al. Angular dose dependence of Matrixx TM and its calibration [ J ]. J Appl Clin Med Phys, 2010, 11(1) :3057. 被引量:1
  • 8Basran PS, Woo MK. An analysis of tolerance levels in IMRT quality assurance procedures [ J ]. Med Phys, 2008, 35 ( 6 ) : 2300-2307. 被引量:1
  • 9Li JG, Yan G, Liu C. Comparison of two commercial detector arrays for IMRT quality assurance [ J ]. J Appl Clin Med Phys, 2009, 10(2) :2942. 被引量:1