期刊文献+

核电厚壁承压容器内壁缺陷TOFD检测优化研究 被引量:5

Optimization of TOFD Detection for Thick-Walled Nuclear Pressure Vessel with Inner Surface Defect
原文传递
导出
摘要 通过分析声波传播特性、检测信号及图像特征,提出20°探头的超声衍射时差法(TOFD)检测模式,并研制20°探头。通过对比5 MHz及2.25 MHz这2种频率下检测效果,进行优化试验。然后通过对不同的内壁模拟缺陷进行检测,制定适合厚壁承压容器内壁缺陷的最佳TOFD检测工艺。试验表明:在核电厂厚壁承压容器内壁缺陷TOFD检测中,20°探头、5 MHz频率是最佳选择;在此检测工艺下,对壁厚120 mm以内的承压容器,能够检出内壁缺陷的最小临界尺寸高为2 mm。 The 20° probe is proposed in order to optimize the TOFD detection. Acoustic wave characteristics, detection signal and image feature of 20° probe are analyzed. The oPtimal frequency is determined by comparing 2.25MHz and 5MHz. Through TOFD detection of artificial inner surface defect, the optimum detections are demonstrated, which are fit for detecting thick-walled nuclear pressure vessel with inner surface defect. The experimental results show that: (a) for thick-walled nuclear pressure vessel with inner surface detect, 20~ probe and 5MHz frequency are optimization. (b) By this optimization detection, artificial inner surface defect with height of 2mm can be effectively distinguished in nuclear pressure vessel with thickness of 120mm.
出处 《核动力工程》 EI CAS CSCD 北大核心 2014年第2期79-83,共5页 Nuclear Power Engineering
基金 中国国电集团公司重大科技专项(GDJJ2012[84])
关键词 TOFD 检测优化 厚壁 承压容器 细晶材质 内壁缺陷 TOFD, Detection optimization, Thick-walled, Pressure vessel, Fine grain material, Inner surface defect
  • 相关文献

参考文献6

二级参考文献38

共引文献60

同被引文献47

引证文献5

二级引证文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部