摘要
血糖浓度是人体生理参数的一项重要生理指标,无创血糖检测仪能够实时检测血糖浓度。设计了一款针对于无创血糖检测仪的温度标定系统,以对称热传递法为基本方法,完成系统的软硬件设计。实验表明,系统能够完成对无创血糖检测仪的标定,标定后无创血糖检测仪温度参数测量误差为0.1摄氏度,满足仪器对相对精度的要求。
Blood glucose concentration is an important physiological indicator of human physiological parameter. Blood glucose concentration can be detected by non-invasive blood glucose monitoring device in real-time.The calibration system with symmetrical heat transfer method as the theoretical basis was designed for the non-invasive blood glucose monitoring.Experimental results showed that the system can be used for the temperature of non-invasive blood glucose detector.The relative error of temperature parameters is 0. 1 ℃ which can meet the requirements of the instrument.
出处
《现代科学仪器》
2013年第5期39-41,共3页
Modern Scientific Instruments
基金
国家863项目
血糖检测与胰岛素注射微系统(2011AA040406)
关键词
温度
标定
血糖
对称热传递法
Temperature
Calibration
Blood Glucose Concentration
Method of Symmetrical Heat Transfer