无针正压密闭输液接头阻菌效果分析 被引量:1

The Performance of the Microbial Barrier of Needleless Positive Pressure Closed Connectors
下载PDF
导出
摘要 目的分析无针正压密闭输液接头的阻菌效果。方法选取三种品牌的无针正压密闭输液接头,分别做微生物侵入试验。结果三种品牌的无针正压密闭输液接头实验结果均呈现为阳性。结论目前市场上有不少的无针正压密闭输液接头产品的阻菌性严格来说还没有达到标准和要求,在结构设计上以及具体的生产工艺上还有很大的改进空间。 Objective Analyzing the performance of microbial barrier of needleless positive pressure closed connectors. Methods Three kinds of brand needleiess positive pressure closed connectors were chosen to do the access of microorganisms test. Results Positive results were detected in the three experimental groups. Conclusions So far, lots of the connectors are not qualified strictly on the market, some improvement is required in the structure design and process.
出处 《中国医疗器械杂志》 CAS 2013年第5期365-366,共2页 Chinese Journal of Medical Instrumentation
关键词 无针正压密闭输液接头 微生物侵入试验 阳性 needleless positive pressure closed connectors, ingress of micro-organisms test, positive result
  • 相关文献

参考文献7

  • 1Hadaway L, Richardson D. Needleless connectors: a primer on terminology[J]. J Infus, 2010, 33(1): 22-31. 被引量:1
  • 2医疗器械质量管理和通用要求标准化委员会,北京国医械华光认证公司.YY/T0316-2008医疗器械风险管理对医疗器械的应用[S]. 被引量:2
  • 3ICU Medical. Products[OL]. http://www.icumed.com. 被引量:1
  • 4国家药典委员会.中华人民共和国药典[M].北京:化学工业出版社,2005.114. 被引量:737
  • 5国家食品药品监督管理局.GB/T14233.2-2005医用输液、输血、注射器具试验方法[S]. 被引量:1
  • 6.消毒技术规范[S].[S].中华人民共和国卫生部,2002,11.. 被引量:1181
  • 7Ye' benes JC, Delgado M, Sauca G, et al. Efficacy of three different valve systems of needle-free closed connectors in avoiding access of microorganisms to endovascular catheters after incorrect handling[J]. Crit Care Med, 2008, 36(9): 2558-2561. 被引量:1

共引文献1917

同被引文献3

引证文献1

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部