摘要
目的探讨双源CT单能谱成像技术在正确评估金属(不锈钢)植入物周围组织CT值中的应用。方法对含有金属植入物的水模体使用双源CT双能量扫描,采用双能量扫描程序,一次性扫描采集2个不同能量(80kV及140kV)的数据,采用单能谱成像技术对所得原始数据进行处理,得到融合后图像。对金属植入物周围的水进行CT值测量和统计分析。结果3种图像(80kVCT图像、140kVCT图像、能谱融合图像)中,能谱融合图像的质量最高,140kVCT图像次之,80kVCT图像最差。用SAS9.1软件分别比较80kVCT图像和140kVCT图像与能谱融合图像沿着金属植入物右侧方向O.1、0.5、1.0、1.5、2.0、2.5、3.0cm的CT值,结果发现它们之间的差异基本上都有统计学意义。结论双源CT单能谱成像技术改善了金属植入物植入后CT检查的图像质量。
Objective To evaluate monoenergetic imaging of dual energy CT in the visualization of metal implants(stainless steel implants). Methods Two different data(80 and 140 kV)ofwater phantom which contained metal implants were collected in one-time scanning using dual-energy scanning sequence. Monoenergetic technique which two different imaging at 80 kV and 140 kV were used for subtraction to removing metal artifacts was applied to generate the new imaging. The CT value of water surrounding metal implants were measured and compared. Results In the three images (80 kV, 140 kV, spectrum image fusion), the latter quality is the highest and the 80 kV scanning images' quality was the worsL The difference between CT values of 80 kV images ( 140 kV images) and energy spectrum images along the right direction 0.1, 0.5, 1, 1.5, 2.0, 2.5, 3.0 cm of the metal basically was significant assessed by the SAS9.1 software. Conclusions Dual energy CT can effectively reduce artifacts of metal implants and will provide high image quality in the area with metal implants.
出处
《国际放射医学核医学杂志》
2013年第4期207-210,共4页
International Journal of Radiation Medicine and Nuclear Medicine