环氧乙烷灭菌生物指示物抗力实验研究 被引量:9

RESEACH ON RESISTANCE OF ETHYLENE OXIDE BIOLOGICAL INDICATOR
原文传递
导出
摘要 目的研究环氧乙烷灭菌生物指示物的抗力,为科学评价环氧乙烷灭菌生物指示剂产品提供科学依据。方法通过专用抗力检测仪,采用细菌定量和定性检测方法,对不同来源的枯草杆菌黑色变种芽孢制作的生物指示物抗力参数进行评价。结果国内外不同企业生产的枯草杆菌黑色变种芽孢生物指示物对环氧乙烷抗力存在差异。6个企业的环氧乙烷生物指示剂的D值分别为2.24、1.64、1.69、1.59、3.19和3.00。不同包装对抗力影响无显著差异,而有机物有明显影响。结论国内外不同企业生产的环氧乙烷生物指示剂抗力不同,需要进行统一标准化和规范化评价。 Objective To study the resistance of ethylene oxide (EO) biological indicator (BI) in order to provide sci entific basis for evaluating. Method The BIs from different manufacturers made of Bacillus atrophacus spores were evalua ting the resistance parameters by using bacterial quantitative and qualitative detection through Biological Indicator Evaluator Resistometer (BIER). Results It was different in the resistance for EO between BIs from different domestic and oversea enterprises. The D value of BIs from 6 manufactures was 2. 24, 1.64, 1.69, 1: 59, 3.19 and 3.00, respectively. There was no significant difference for BIs' resistance by using different packaging except organics. Conclusion The resistance of BIs from different domestic and oversea enterprises was different, which should be evaluated in standardization mode.
出处 《中国消毒学杂志》 CAS 北大核心 2013年第5期402-404,共3页 Chinese Journal of Disinfection
关键词 环氧乙烷 生物指示物 枯草杆菌黑色变种 D值 抗力 ethylene oxide biological indicator Bacillus atrophaeus D value resistance
  • 相关文献

参考文献5

  • 1黄靖雄,周适明.工业环氧乙烷灭菌的验证和常规监测[J].中国医疗器械信息,2010,16(8):33-36. 被引量:5
  • 2Caputo RA,Rohn KJ.The effects of EtO sterilization variables onBI performance[J].Med Device Diagn Ind,1982,4(7):37. 被引量:1
  • 3Shintani H,Sasaki K,Kajiwara Y,et al.Validation of D-valueby different SCD culture medium manufacturer and/or differentSCD culture medium constituent[J].PDA J Pharm Sci Technol,2000,54(1):6. 被引量:1
  • 4Gregg AM,Clark WH.Interpreting and Understanding MicrobialData in Validation of Ethylene Oxide Sterilization Processes[J].Biomed Instrum Technol,2005,39(6):466. 被引量:1
  • 5Sterilization of health care products-Biological indicators[S].ISO11138-2006,2006. 被引量:1

共引文献4

同被引文献36

  • 1张霞,周雯,王有森.环氧乙烷在灭菌物品中残留量测定及毒性研究进展[J].中国消毒学杂志,2005,22(2):217-218. 被引量:56
  • 2沈菊华.国内外环氧乙烷生产技术及市场分析[J].化工技术经济,2005,23(11):20-22. 被引量:14
  • 3田桂登,陈春丽,田雍,谭家琼.环氧乙烷灭菌过程控制[J].中国消毒学杂志,2007,24(1):95-96. 被引量:2
  • 4Fritze, D., and Pukall, R.2001.Reclassification of bioindicator Ba- cillus subtilis DSM 675 and Bacillus subtilis DSM 2277 as Bacillus atropbaeus[J].Int J Syst Evol Micr,2001,51: 35-37. 被引量:1
  • 5Nakamura, L K.Taxonomic relationship of black-pigmented Bacil- lus subtilis strains and a proposal for Bacillus atrophaeus sp. nov[J].Intemational Journal of Systematic Bacteriology, 1989,39 (3): 295 - 300. 被引量:1
  • 6Lewis,D.L.,M.Arens.Resistance of microorganisms to disinfection in dental and medical services[J].Nat.Med.1995,1:956- 958. 被引量:1
  • 7Dancer, S.J.Importance of the environment in meticillin-resistant Staphylococcus aureus acquisition: the case for hospital cleaning [J].Lancet Infect.Dis, 2008,8 : 101 - 113. 被引量:1
  • 8The United States Pharmacopoeial Convention.USP35-NF30.Unit- ed States Pharmacopoeia[S].THE United States: United States Pharmacopoeia Convection Rockyille,2011:2 357-2 361. 被引量:1
  • 9EDQM.European Pharmacopoeia(Seven Edition)[S].France:Europe- an Pharmacopoeia Commission,2010. 被引量:1
  • 10中华人民共和国卫生部.GB/T15981~1995消毒与灭菌效果的评价方法与标准[S].北京:中国标准出版社,1995:257. 被引量:2

引证文献9

二级引证文献39

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部