基于压差式流量传感器的便携式肺功能仪的重复性研究

Reproducibility of a Portable Spirometer Based on Differential Pressure Sensor
下载PDF
导出
摘要 目的评估基于压差式流量传感器的便携式肺功能仪的重复性,包括操作者间重复性及日间重复性。方法两位操作者分别对22名健康志愿者进行肺通气功能测定,24h后由第一位操作者再进行一次肺通气功能测定。结果两位操作者间及日间各指标的ICC均大于0.75,且差异无统计学意义(P>0.05);Bland-Altman图示所有数据中只有1-2个数据位于一致性范围以外。结论基于压差式流量传感器的便携式肺功能仪重复性很好,可用于临床肺功能检测。 Objective Assessing the reproducibility of a portable spirometer, including repoducibility of inter-observer and day- today. Methods Lung ventilation function was performed in 22 healthy volunteers by two observers on the same day and repeated by the first observer after 24h. Results The inter-observer and day-to-day intra-class correlation coefficients are all higher than 0.75. There are no significant difference between each other. Bland-Altman chart shows good limits of agreement between inter-observer and day-to-day, only scattered data are outside of the limits of agreement. Conclusions The portable spirometer shows good inter-observer and day-to-day reproducibility, and can be used for testing lung function in clinical.
出处 《中国医疗器械杂志》 CAS 2013年第1期30-32,48,共4页 Chinese Journal of Medical Instrumentation
基金 中国科学院知识创新工程领域前沿重点项目(编号0923A11291)
关键词 肺功能仪 肺功能测试 重复性 组间重复性 日间重复性 lung function apparatus, pulmonary function testing, reproducibility, inter-observer reproducibility, day-to-dayreproducibility
  • 相关文献

参考文献9

二级参考文献28

  • 1徐勇勇 赫元涛.测量手段的效度和信度[A].方积乾 主编.医学统计学与电脑实验:第2版[C].上海:上海科学技术出版社,2001.238-251. 被引量:1
  • 2Murakami T, Takei K, Ueno M, et al. Aortic reservoir function after arterial switch operation in elementary school-aged children [ J ]. Circulation Journal, 2008,72 ( 8 ) : 1291-1295. 被引量:1
  • 3Brooks BA, Molyneaux LM, Yue DK. Ahgmentation of central arterial pressure in Type 2 diabetes [ J ]. Diabetic Medicine ,2001,18 (5) :374-380. 被引量:1
  • 4Buckberg GD, Fixler DE, Archie JP, et al. Experimental subendocardial ischaemia in dogs with normal coronary arteries[J]. Circ Res,1972,30( 1 ) :67-81. 被引量:1
  • 5BuckbergGD, Towers B, Paglia DE, et al. Subendocardial ischemia after cardiopulmonary bypass [ J ]. JThorac Cardiovasc Surg, 1972,64 (5) :669-685. 被引量:1
  • 6Nichols WW, O' Rourke MF. McDonald' s Blood Flow in Arteries: Theoretical, Experimental and Clinical Princi ples [ M ]. 5 th. London : Edward Arnold,2005:32-42. 被引量:1
  • 7Safar ME, Levy BI. Current perspectives on arterial stiffness and pulse pressure in hypertension and cardiovascular disease [ J ]. Circulation,2003,107 (22) : 2864-2869. 被引量:1
  • 8Mitchell GF. Arterial stiffness and wave reflection in hy pertension:Pathophysiologic and therapeutic implications [J]. Curt Hypertens Rep,2004,6(6) :436-441. 被引量:1
  • 9Saito M, Kasuya A. Relationship between the subendocar dial viability ratio and risk factors for ischemic heart dis ease[ J ]. Sangyo Eiseigaku Zasshi, 2003,45 ( 3 ) : 114- 119. 被引量:1
  • 10Fokkema DS, van Teeffelen JWGE, Dekker S, et al. Diastolic time fraction as a determinant of subendocardial perfusion[ J ]. Am J Physiol Heart Circ Physiol, 2005, 288 ( 5 ) :2450-2456. 被引量:1

共引文献46

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部