期刊文献+

热管式黑体辐射源研制与评价技术研究 被引量:5

Research of Heat Pipe Blackbody Designing and Assessing Technology
下载PDF
导出
摘要 为了解决辐射温度计的检定、校准过程中黑体辐射源溯源不完善的问题,设计制作了温度范围覆盖200~1100℃的系列热管式黑体辐射源,研究了黑体辐射源的评价方法,测试了其温度分布性能指标,并使用蒙特卡洛法对其有效发射率进行了计算。测试计算结果表明,所制作的热管式黑体辐射源靶面均匀性不大于0.2℃,有效发射率高于0.9995。 To solve the problem of incomplete tracing of blackbody in verification and calibration of radiation thermometers, this paper de- signed a series of heat pipe blackbodies which temperature range covered 200 - 1100 ℃, and researched blackbody assessing technology. The temperature technical indexes of heat pipe blackbodies have been tested, and their effective emissivities have been calculated using the Monte - Carlo method. The result shows that the heat pipe blackbody target surface uniformity is less than 0. 2 ℃, and the effective emissivity is greater than 0. 9995.
作者 张学聪
出处 《计测技术》 2012年第4期30-32,44,共4页 Metrology & Measurement Technology
基金 航空科学基金资助项目(20102244003)
关键词 热管 黑体 发射率 蒙特卡洛法 heat pipe blackbody emissivity Monte-Carlo mthod
  • 相关文献

参考文献9

二级参考文献26

共引文献73

同被引文献43

  • 1闫小克,吕卓凡,马重芳,李杰,段宇宁,Xiao-ke Zhuo-fan Chong-fang Yu-ning.高精度钠热管固定点炉及其等温特性[J].计量学报,2009(6). 被引量:7
  • 2黄东涛,陆家钦,段宇宁,原遵东,赵琪.黑体空腔发射率计算的蒙特卡罗模型[J].清华大学学报(自然科学版),1997,37(2):28-31. 被引量:17
  • 3SAUNDERS P, FISCHER J, SADLI M, et al. Uncer- tainty budgets for calibration of radiation thermometers below the silver point [ J ]. International Journal of Ther- mophysics, 2008, 29(3) : 1066-1083. 被引量:1
  • 4郝小鹏,原遵东,王景辉,等.1.6微米近红外辐射温度计研制[J].仪器仪表学报,2010,31(8):231-235. 被引量:1
  • 5原遵东,王景辉,郝小鹏,等.变温黑体辐射源溯源方法探讨[J].计量学报,2012,33(z2):1-3. 被引量:1
  • 6ZHUY, MA H, WANG R. -50 to +150 ℃ heat pipe blackbody source for radiation thermometer calibration[ A ]. Temperature: Its Measurement and Control in Science and Industry, 1982, (1) : 559-565. 被引量:1
  • 7HOLLANDT J, FRIEDRICH R, GUTSCHWAGER B, et al. High-accuracy radiation thermometry at the national metrology institute of germany, the PTB [ J ]. High Tem- peratures, High Pressures, 2003, 35(4): 379-415. 被引量:1
  • 8NOORMA M, MEKHONTSEV S, KHROMCHENKO V, et al. Water heat pipe blackbody as a reference spectral radiance source between 50 ℃ and 250 ℃ [ C]. Defense and Security Symposium. International Society for Optics and Photonics, 2006: 620502-620502-9. 被引量:1
  • 9HILL K D, WOODS D J. Characterizing the NRC black- body sources for radiation thermometry from 150 ℃ to 962 ℃ [ J ]. International Journal of Thermophysics, 2009, 30( 1 ) : 105-123. 被引量:1
  • 10MACHIN G, CHU B, High quality blackbody sources for infrared thermometry and thermography between -40 ℃ and 1 000 ℃ [ J ]. Imaging Science Journal, 2000, 48 (1) : 15-22. 被引量:1

引证文献5