期刊文献+

国产与进口ICP-OES仪器检测遥控按键中Cd和Pb的对比研究 被引量:1

Study on the Determination of Cadmium and Plumbum in Remote Control Button by Domestic and Imported ICP-OES
下载PDF
导出
摘要 针对遥控按键中的镉(Cd)和铅(Pb),使用HNO3和H2O2经微波溶解过滤后,分别使用国产和进口的电感耦合等离子体发射光谱仪(ICP-OES)进行检测。结果表明,两种仪器的检测结果相近,白色和蓝色按键中镉的检出结果相差不大于4ppm,而橙色按键中铅的检出结果相差小于6.2%。通过比对进口与国产ICP-OES的检测情况,简要地说明了使用ICP-OES检测塑胶遥控按键中Cd和Pb的一般方法及步骤,并为相关检测单位和人员根据不同需求选择不同仪器提供了参考。 Cd and Pb in the remote control button are dissolved by HNO3 and H2O2 with microwave instrument,then the solutions are filtered and detected by domestic and imported Inductively Coupled Plasma Emission Spectrometer(ICP) respectively.The results show that the two instruments have the similar detection conclusion,The detection deviation of Cd are less than 3ppm between two instruments in white and blue buttons,while the detection deviation of Pb is less than 6.2% in orange button.For sum up,we briefly describes the detection of Cd and Pb by ICP-OES in plastic remote control button of RoHS index,and provides a reference of selecting corresponding instruments according the different needs for related detection unit.
出处 《现代科学仪器》 2012年第4期95-98,共4页 Modern Scientific Instruments
基金 江西省自然科学基金(2010GZH0002)项目资助
关键词 江苏天瑞仪器ICP-2000B PerkinElmer OES-7300DV ROHS Jiangsu Skyray Instrument ICP-2000B Perkinelmer OES-7300DV RoHS Cadmium Plumbum
  • 相关文献

参考文献5

二级参考文献38

  • 1汪禄祥,刘家富,董宝生,严红梅.微波消解氢化物-原子荧光法同时测定土壤中的砷和汞[J].西南农业学报,2006,19(2):211-213. 被引量:25
  • 2刘翠平,夏碧琪.ICP-AES测定塑料中的镉的不确定度评定[J].光谱实验室,2007,24(2):73-77. 被引量:10
  • 3朱明华.仪器分析[M].北京:高等教育出版社,2005:196-224. 被引量:4
  • 4ISO 11885-1996,Water quality-determination of 33 elements by inductively coupled plasma atomic emission spectrometry[S]. 被引量:1
  • 5ISO 17294-1-2004,Water quality-Application of induetively coupled plasma mass spectrometry (ICP-MS) for the determination of elements-Part1:General guidelines IS]. 被引量:1
  • 6JIS K 0116-2003,General rules for atomic emission spectrometry[S]. 被引量:1
  • 7JIS K 0133-2007,General roles for high frequency plasma mass spectrometry[S]. 被引量:1
  • 8H. Liu, A. Montaser, Phase-doppler diagnostic stud- ies of primary and tertiary aerosols produced by a high- efficiency nebulizer, Anal. Chem. 66 (1994) 3233-3242. 被引量:1
  • 9J.L. Todol i, V. Hemandis, A. Canals, 3.M. Mermet, Comparison of characteristics and limits of detection of pneumatic micro-nebulizers and a conventional nebulizer operating at low uptake rates in ICP-AES, I. Anal. At. Spectrom. 14 (1999) 1289-1295. 被引量:1
  • 10S.E. Maestre, J.L. Todol i, J.M. Mermet, Evaluation of several pneumatic micronebulizers with different designs for use in ICP-AES and ICP-MS. Future directions for further improvement, Anal. Bioanal. Chem. 379 (2004) 888-899. 被引量:1