不同参数设置对植入型单腔心律转复除颤器不恰当放电的影响 被引量:1

Effect of Different Detection Algorithms on Inappropriate Shocks Caused by Single Chamber Implantable Cardioverter Defibrillators
下载PDF
导出
摘要 目的 探讨单腔植入型心律转复除颤器(ICD) 不同识别参数设置对不恰当放电的影响.方法 A组入选25例2007年至2008年植入单腔ICD的患者,术后经验设置室性心律失常与室上性心动过速鉴别参数.B组入选25例2009年至2010年植入单腔ICD的患者,术后开启稳定性、突发性和形态学标准.每例随访1年.观察两组室上性心动过速正确识别率和不恰当放电率.结果 与A组相比,B组对室上性心动过速正确识别率较高(68.0%、28.0%),室上性心动过速放电率较低(24.0%、76.0%),不恰当放电所致住院率较低(4.0%、20.0%).A组不恰当放电引发室性心动过速2例.结论 术后启用三项鉴别参数能减少ICD不恰当放电. Objective To explore the effect of different detection algorithms on inappropriate shocks caused by single chamber implantable cardioverter defibrillators (ICD). Methods 25 patients implanted with single chamber ICD during 2007-2008 were included in group A and another 25 patients implanted with single chamber ICD during 2009-2010 were included in group B. The parameters identifying ventricular and supraventricular tachycardia(SVT) were set based on per- sonal experiences in group A. While the criteria such as stability, sudden-onset and morphology were implemented after the procedure in group B. All the patients were followed up for a year. The diagnosis accuracy of SVT and inappropriate shocks were analyzed. Results The percentage of correct diagnoses of SVT was higher in group B than in group A (68.0% vs.28.0%). Inappropriate shocks and hospitalization for inappropriate shocks were less in group B (20.0% vs. 84.0%) than in group A (24.0% vs.76.0%). Ventricular tachycardia was induced by inappropriate shocks in 2 cases of group A. Conclusion The application of stability, sudden-onset and morphology after procedure may reduce inappro- priate shocks caused by ICD.
出处 《心电与循环》 2012年第3期155-156,184,共3页 Journal of Electrocardiology and Circulation
基金 温州市科技局课题(Y20110156)
关键词 植入型心律转复除颤器 不恰当放电 室上性心动过速 Implantable cardioverter defibrillators Inappropriate shocks Supraventricular tachcardia
  • 相关文献

参考文献7

  • 1沈法荣.植入式心脏复律除颤器患者放电后管理[J].心血管病学进展,2011,32(5):614-617. 被引量:5
  • 2Gollob MH,Seger JJ. Current status of the implantable cardioverter-defibrillator[J].Chest,2001.1210-1221. 被引量:1
  • 3Jackson LR,Daubert JP,Thomas KL. Expanding the benefits of implantable cardioverter-defibrillator therapy:"is less more"[J].Progress in Cardiovascular Diseases,2012.372-378. 被引量:1
  • 4Simmers TA. Remote care:bidding our ICD patients a fond farewell[J].Neth Heart J,2012.71-76. 被引量:1
  • 5Theuns DA,Rivero-Ayerza M,Goedhart DM. Evaluation of morphology discrimination for ventricular tachycardia diagnosis in implantable cardioverter-defibrillators[J].Heart Rhythm,2006.1332-1338. 被引量:1
  • 6Mainigi SK,Almuti K. Usefulness of radiofrequency ablation of supraventricular tachycardia to decrease inappropriate shocks from implantable cardioverter-defibrillators[M].American Journal of Cardiology,2012.15,109∶231-237. 被引量:1
  • 7Kuhlkamp V,D(o)rnberger V,Mewis C. Clinical experience with the new detection algorithms for atrial fibrillation of a defibrillator with dual chamber sensing and pacing[J].Journal of Cardiovascular Electrophysiology,1999.905-915. 被引量:1

二级参考文献20

  • 1Braunschweig F, Boriani G, Bauer A, et al. Management of patients receiving implantable cardiac defibrillator shocks [ J ]. Europace, 2010, 12 ( 12 ) : 1673- 1690. 被引量:1
  • 2Pinski SL, Yao Q, Epstein AE, et al. Determinants of outcome in patients with sustained ventricular tachyarrhythmias: the Antiarrhythmics Versus Imp|antable Defibrillators (AVID) study registry [ J ]. Am Heart J, 2000, 139 ( 5 ) : 804- 813,. 被引量:1
  • 3Cleland JG, Daubert JC, Erdmann E, et al. The effect of cardiac resynchroni- zation on morbidity and mortality in heart failure [ J ]. N Engl J Med, 2005, 352 : 1539-1549. 被引量:1
  • 4Poole JE,Johnson GW, Hellkamp AS, et al. Prognostic importance of defibrilla- tor[ J ]. NEJM, 2008, 359 : 1009-1017,. 被引量:1
  • 5Clements PA. Hazards of performing chest compressions in collapsed patients with internal eardioverter defibrillators[ J]. Emerg Med J ,2003,20:379-380. 被引量:1
  • 6Israel CW, Barold SS. Electrical storm in patients with an implanted defibrillator: a matter of definition [ J ], Ann Noninvasive Electroeardiol, 2007,12 : 375- 382. 被引量:1
  • 7Wathen M, Sweeney M, DeGroot P. Shock reduction using antitachycardia pa- cing for spontaneous rapid ventricular tachycardia in patients with coronary artery disease [J]. Circulation,2001,104: 796-801. 被引量:1
  • 8Wathen MS, DeGroot PJ, Sweeney MO, et al. Prospective randomized multi- center trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter defibrina- tom : PainFREE RxlI Trial Results [ J ]. Circulation,2004,110:2591-2596. 被引量:1
  • 9Schwab JO. EnTrust Clinical Study Results. HRS ( May 2005 ). 被引量:1
  • 10Poole JE. Analysis of ICD shock electrograms in the SCD-HeFT trial [ R ]. Heart Rhythm Society Conference, 2004. Poster. 被引量:1

共引文献4

同被引文献16

  • 1Chen S,Yin Y,Krucoff M W. Effect of cardiac resynchronization therapy and implantable cardioverter defibrillator on quality of life in patients with heart failure:a meta-analysis[J].EUROPACE,2012,(11):1602-1607. 被引量:1
  • 2Jordaens L J,Mekel J M. Electrical storm in the ICD era[J].EUROPACE,2005,(02):181-183. 被引量:1
  • 3Gatzoulis K A,Andrikopoulos G K,Apostolopoulos T. Electrical storm is an in-dependent predictor of adverse long-term outcome in the era of implantable defibril ator therapy[J].EUROPACE,2005,(02):184-192. 被引量:1
  • 4Streitner F,Kuschyk J,Veltmann C. Predictors of electrical storm recurrences in patients with implantable cardioverter-defibrill-ators[J].EUROPACE,2011,(05):668-674. 被引量:1
  • 5Epstein A E,DiMarco J P,Ellenbogen K A. 2012 ACCF/AHA/HRS focused update incorporated into the ACCF/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities:a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines and the Heart Rhythm Society[J].CIRCULATION,2013,(03):e283-e352. 被引量:1
  • 6Hohnloser S H,Al-Khalidi H R,Pratt C M. Electrical storm in patients with an implantable defibrillator:incidence,features,and preventive therapy:insights from a randomized trial[J].EUROPEAN HEART JOURNAL,2006,(24):3027-3032. 被引量:1
  • 7Arya A,Haghjoo M,Dehghani M R. Prevalence and predictors of electrical storm in patients with implantable cardioverter-defibri-llator[J].American Journal of Cardiology,2006,(03):389-392. 被引量:1
  • 8Brigadeau F,Kouakam C,Klug D. Clinical predictors and prog-nostic significance of electrical storm in patients with implantable cardioverter defibrillators[J].EUROPEAN HEART JOURNAL,2006,(06):700-707. 被引量:1
  • 9Israel C W,Barold S S. Electrical storm in patients with an implant-ed defibrillator:a matter of definition[J].Ann Noninvasive Electro-cardiol,2007,(04):375-382. 被引量:1
  • 10Verma A,Kilicaslan F,Marrouche N F. Prevalence,predictors,and mortality significance of the causative arrhythmia in patients with electrical storm[J].JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY,2004,(11):1265-1270. 被引量:1

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部