期刊文献+

原子吸收光谱法测定婴幼儿奶粉中铁含量不确定度的评定 被引量:3

Uncertainty Evaluation of Determination of Iron in Milk Powder for Infants by AAS
原文传递
导出
摘要 建立了原子吸收光谱法测定婴幼儿奶粉中铁含量不确定度评定的数学模型,分析了检测过程中不确定度来源并计算了各不确定度分量,得出合成标准不确定度和扩展不确定度。在婴幼儿奶粉中铁含量的测定中,测量值为8.156mg/100g,其扩展不确定度为0.11mg/100g,k=2。影响铁含量测量结果不确定度的最大因素是校准曲线拟合引入的不确定度。 The mathtic mode of uncertainty evaluation of the content of iron in milk powder for infants by atomic absorption spectrometry(AAS) was established,and the sources of uncertainty were analyzed,and the each component of uncertainty was calculated,and the combined standard and expanded uncertainty were obtained.The content of iron in milk powder for infants was 8.156mg/100g,and the expanded uncertainty was 0.11mg/100g,k=2.The influence of maximum factor on the uncertainty of the determination results for iron content was calibration curve fitting uncertainty.
作者 张荷丽
出处 《光谱实验室》 CAS CSCD 2012年第3期1501-1505,共5页 Chinese Journal of Spectroscopy Laboratory
基金 河南省高等学校青年骨干教师资助计划项目(2009GGJS-128)
关键词 奶粉 铁含量 原子吸收光谱法 不确定度 Milk Powder Iron Content AAS Uncertainty
  • 相关文献

参考文献3

  • 1国家质量技术监督局.测量不确定度评定与表示[S].JJF 1059-1999.北京:中国计量出版社,1999. 被引量:1
  • 2中华人民共和国卫生部.食品安全国家标准婴幼儿食品和乳品中钙、铁、辞、钠、钾、镁、铜和锰的测定[S].GB5413.21-2010.北京:中国标准出版社,2010. 被引量:1
  • 3国家质量监督检疫检验总局.常用玻璃量具[S].JJG 196-2006.北京:中国计量出版社,2006. 被引量:1

同被引文献24

引证文献3

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部