摘要 目的:测量液电型碎石机总的放电噪声以及开关噪声和电极噪声的大小。方法:选用最常用的碎石机高压电容、开放式火花隙开关和放电电极,借助隔音箱,在不同电压时,用声级计测出操作医生耳朵部位的总放电噪声、开关噪声和电极噪声。结果:在最低放电电压5 kV时,碎石机总放电噪声、开关产生的噪声和电极产生的噪声分别是108.5、107.7、82 dB;在最高放电电压11 kV时,碎石机总放电噪声、开关产生的噪声和电极产生的噪声分别是118、116.7、93.2 dB,高于国家相关标准规定的90 dB。结论:医生和患者应当采取听力保护措施。 Objective To measure and analyze the spark gap noise,electrode noise and their comprehensive noise in the electrohydraulic extracorporeal shock wave lithotripsy(ESWL).Methods Using the usual high voltage pulse capacitor,open-type spark gap switch and electrodes,the noises levels at doctor's ear position in different treating voltages were measured by use of sound level meter and sound shielding box.The readings were taken and then averaged.Results The comprehensive noise,spark gap noise and electrode noise at the minimum treating voltage 5 kV were 108.5 dB,107.7 dB and 82 dB, respectively;the above noises at the maximum treating voltage 11 kV were 118 dB,116.7 dB and 93.2 dB,respectively.All comprehensive noise readings were over 90 dB,which was the occupational safety and health administration(AOSHA) standard.Conclusion Hearing protection to the doctors and patients is considered necessary.
作者 马世宏
出处 《医疗卫生装备》 CAS 2012年第5期116-117,共2页 Chinese Medical Equipment Journal
关键词 体外碎石机 火花隙开关 声级计 噪声 electrohydraulic extracorporeal shock wave lithotripsy(ESWL) spark gap switch sound level meter noise
  • 相关文献

参考文献9

二级参考文献18

  • 1马世宏,张春野.碎石机中电极附极杆套管破裂损坏的原因分析[J].电工电能新技术,1995,14(2):38-42. 被引量:2
  • 2马世宏.防松脱传输线:中国,ZL2009201460505[P].2009-02-14. 被引量:2
  • 3马世宏.几种国外体外冲击波碎石机的比较.国外医学 泌尿系统分册,1994,(14):41-42. 被引量:1
  • 4马世宏.碎石机电极:中国.ZL200920146049.2[P].2009-02-14. 被引量:1
  • 5杜致远.自由振膜式自聚焦冲击波发生器:中国,ZL200520069071.3[P].2005-02-06. 被引量:1
  • 6马世宏.可调节长寿命火花隙开关:中国,ZL200820136438.2[P].2008-02-13. 被引量:2
  • 7Kuender D.Yang.Extraeorporeal shockwave treatment for chronic diabetic foot ulcers[J].Journal of Surgical Research,2009,152(1):96-103. 被引量:1
  • 8陈忠兴.冲击波发生装置及冲击波减肥仪:中国,ZL200720047654.52[P].2008-02-04. 被引量:1
  • 91,严忠礼,等.水中瞬态压力测量技术与应用.力学未来15年国际学术讨论会论文集,第二卷,1986 被引量:1
  • 102,体外冲击波碎石机通用技术条件.1991 被引量:1

共引文献14

同被引文献6

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部