摘要
本文论述了使用毛细管柱和FID检测器 ,采用气相色谱程序升温法对柴油(0#)进行分析 ,与磺化法相比 ,两种方法的标准偏差分别为0.026和1.19,变异系数分别为0.075 %和4.53% ,回收率分别为98.0%~102.8%和114.2%~134.4 %。
This report introduced GC method with capillary column and FID detecter to analysis petroleum oils,using temperature program from 85℃ to 210℃ and tributyl phosphate as interal standard. Comparation with GC and sulphonating ,the results showed that the variation coeffcients were 0.075% and 4.53%,the standard deviation were 0.026 and 1.19, the recoveries were 98.0%~102.8% and 114.2%~134.4% respectively.
出处
《农药科学与管理》
CAS
2000年第2期2-3,共2页
Pesticide Science and Administration