可视吸痰系统的离体实验 被引量:4

Experiment of visual suction system in vitro
下载PDF
导出
摘要 目的探讨采用可视吸痰系统进行离体肺、气管内吸痰的可行性及有效性。方法在离体猪肺、气管的不同部位注入浓度为1.5%、3.0%的模拟痰液,吸痰的进入通道分别为咽喉插入吸痰、气管插管吸痰及气管切开吸痰,根据吸痰方式,将实验猪分为传统吸痰组(采用传统吸痰系统进行标准吸痰操作,n=40)及可视吸痰组(采用可视吸痰系统进行标准吸痰操作,n=40),吸痰时间为15s,设定吸痰负压为-200mm Hg。对两组的吸痰量进行比较。结果用1.5%浓度的模拟痰液进行试验,可视吸痰组采用咽喉插入吸痰、气管插管吸痰及气管切开吸痰的吸痰量分别为(6.91±0.49)、(7.27±0.71)及(6.59±0.88)g,传统吸痰组的吸痰量分别为(5.28±0.64)、(6.39±0.91)及(5.98±1.38)g;用3.0%模拟痰液进行试验,可视吸痰组采用咽喉插入吸痰、气管插管吸痰及气管切开吸痰的吸痰量分别为(6.37±1.15)、(7.98±0.97)及(7.35±1.79)g,传统吸痰组的吸痰量分别为(5.64±1.29)、(6.19±1.05)及(6.90±1.25)g;可视吸痰组与传统吸痰组的吸痰量比较,差异均有统计学意义(P<0.05)。结论可视吸痰系统的吸痰效率比传统吸痰系统高,且更易操作,具有临床可行性。 Objective To explore the feasibility and effectiveness of visual sputum suction used in sputum suction in lung and trachea in vitro.Methods Concentration of 1.5% and 3.0% of simulated sputum were injected into different parts of pig lung and trachea in vitro.The pathways of sputum suction were through throat insertion,trachea cannula and tracheotomy,respectively.The experimental pigs were divided into conventional sputum suction group(conventional sputum suction system used to conduct standard suction,n=40) and visual sputum suction group(visual sputum suction system used to conduct standard suction,n=40).Duration of suction was 15 secends,and the suction negative pressure was set to be-200 mm Hg.The suction capacities of the two groups were compared.Results Experiment with 1.5% simulated sputum,the suction capacities of methods of throat insertion,trachea cannula and tracheotomy in visual sputum suction group were(6.91±0.49),(7.27±0.71) and(6.59±0.88)g,while those in conventional sputum suction group were(5.28±0.64),(6.39±0.91) and(5.98±1.38)g.Experiment with 3.0% simulated sputum,the suction capacities of methods of throat insertion,trachea cannula and tracheotomy in visual sputum suction group were(6.37±1.15),(7.98±0.97) and(7.35±1.79)g,while those in conventional sputum suction group were(5.64±1.29),(6.19±1.05) and(6.90±1.25)g.Compared the sputum capacities between visual sputum suction group and conventional sputum suction group,the differences were all statistically significant(P0.05).Conclusion Sputum suction efficiency of visual sputum suction system is higher than that of conventional sputum suction system,with easier operation and clinical feasibility.
出处 《重庆医学》 CAS CSCD 北大核心 2012年第9期869-871,共3页 Chongqing medicine
基金 重庆市科委重点攻关基金资助项目(CSTC 2009AB5218)
关键词 纤维光学 动物 实验 可视吸痰 sputum fiber optics animals laboratory visual sputum suction
  • 相关文献

参考文献12

  • 1尹利华,王建荣,张利岩,张健鹏.密闭式吸痰研究进展[J].南方护理学报,2005,12(4):13-15. 被引量:73
  • 2Barnes CA,Kirchhoff KT.Minimizing hypoxemia due toendotracheal suctioning:a review of the literature[J].Heart Lung,1986,15(2):164-176. 被引量:1
  • 3Johnson KL,Kearney PA,Johnson SB,et al.Closed ver-sus open endotracheal suctioning:costs and physiologicConsequences[J].Crit Care Med,1994,22(4):658-666. 被引量:1
  • 4Reissmann H,Bhm SH,Suárez-Sipmann F,et al.Suctio-ning through a double-lumen endotracheal tube helps toprevent alveolar collapse and to preserve ventilation[J].Intensive Care Med,2005,31(3):431-440. 被引量:1
  • 5Brochard L,Mion G,Isabey D,et al.Constant-flow insuf-flation prevents arterial Oxygen desaturation during endo-tracheal suctioning[J].Am Rev Respir Dis,1991,144(2):395-400. 被引量:1
  • 6Maggiore SM,Lellouche F,Pigeot J,et al.Prevention ofendotracheal suctioning-induced alveolar derecruitment inacute lung injury[J].Am J Respir Crit Care Med,2003,167(9):1215-1224. 被引量:1
  • 7Shah S,Fung K,Brim S,et al.An in vitro evaluation ofthe effectiveness of endotracheal suction catheters[J].Chest,2005,128(5):3699-3704. 被引量:1
  • 8Pedersen CM,Rosendahl-Nielsen M,Hjermind J,et al.Endotracheal suctioning of the adult intubated pa-tient——what is the evidence[J].Intensive Crit CareNurs,2009,25(1):21-30. 被引量:1
  • 9刘宏见.机械通气患者气管内开放式与密闭式吸痰效果比较[J].当代护士(中旬刊),2008,15(2):84-85. 被引量:13
  • 10Liu GY,Lu JS,Shen KW,et al.Fiberoptic ductoscopycombined with cytology testing in the patients of sponta-neous nipple discharge[J].Breast Cancer Res Treat,2008,108(2):271-277. 被引量:1

二级参考文献26

  • 1杜斌.呼吸机相关性肺炎[J].中华医学杂志,2002,82(2):141-144. 被引量:200
  • 2杨霞,范学朋.密闭式吸痰器在呼吸机治疗中的应用与护理[J].中华护理杂志,2005,40(3):196-198. 被引量:100
  • 3卢中秋,李玉苹,李惠萍,黄唯佳,李景荣,李章平,吴进泽.床旁纤维支气管镜技术在急救中的应用[J].世界急危重病医学杂志,2005,2(4):778-780. 被引量:3
  • 4Wainwright SP, Gould D. Endotracheal suction: an example of the problems of relevance and rigour in clinical research[ J ]. Journal of Clinical Nursing,1996,5(6):389-398. 被引量:1
  • 5A. Topeli, A. Harmanci, Y. Cetinkaya, et al. Comparison of the effect of closed versus open endotracheal suction systems on the development of ventilator-associated pneumonia[J].Journal of Hospital Infection, 2004,58:14-19. 被引量:1
  • 6C. C. Freytag, F. L. Thies. Prolonged application of closed in-line suction catheters increases microbial colonization of the lower respiratory tract and bacterial growth on catheters surface[J]. Infection,2003,31(1):31-37. 被引量:1
  • 7Salvatore M. Maggiore, Francois Lellouche, Jerome Pigeot, et al. Prevention of endotracheal suction-induced alveolar derecruitment in acute lung injury[J]. American Journal of Respiratory and Grit Care Medieine,2003,167(9):1215-1224. 被引量:1
  • 8Maurizio Cereda, Federico Villa, Enrico Colombo, et al.Closed system endaotracheal suctioning maintains lung volume during volume-controlled mechanical ventilation[J].Intensive Care Med,2001,27:64-654. 被引量:1
  • 9Angela Kalyn, Sandra Feuerstake, Bosco Paes, et al. Closed suction of intubated neonates maintains better physiologic stability: a randomized trial[J]. Journal of Perinatoligy,2003, 23:218-222. 被引量:1
  • 10Leandro Cordero, MD, Mercedes Sananes, et al. Comparison of a closed (Trach Care MAC) with an open endotracheal suction system in small premature infants[ J ]. Journal of Perinatology,2000, 3:151-156. 被引量:1

共引文献109

同被引文献42

  • 1Martin-Loeehes I, Artigas A, Gordo F, et al. Current status of fibreoptic bronchoscopy in intensive care medicine. Med Intensi- va, 2012,36(9):644-649. 被引量:1
  • 2Nair A, Godoy MC, Holden EL, et al. Multidetector CT and postprocessing in planning and assisting in minimally invasive bronchoscopic airway interventions. Radiographics, 2012,32(5): 201-232. 被引量:1
  • 3Gibbs JD, Graham MW, Higgins WE. 3D MDCT-based system for planning peripheral bronchoscopic procedures. Comput Biol Med, 2009,39(3) :266-279. 被引量:1
  • 4Lv J, Wu J, Guo R, et al. Laboratory test of a visual sputum suctioning system. Respir Care, 2013,58(10):1637-1642. 被引量:1
  • 5Kotloff RM, Thabut G. Lung transplantation. Am J Respir Crit Care Med, 2011,184(2) :159-171. 被引量:1
  • 6Cappuzzo F, Bartolini S, Crino L. Emerging drugs for non-small cell lung cancer. Expert Opin Emerg Drugs, 2003,8(1) :179-192. 被引量:1
  • 7Radzikowska E, Glaz P, Roszkowski K. Lung cancer in women : Age, smoking, histology, performance status, stage, initial treatment and survival. Population-based study of 20 561 cases. Ann Oncol, 2002,13(7) :1087-1093. 被引量:1
  • 8Ryu JH, Myers JL, Swensen SJ. Bronchiolar disorders. Am J Respir Crit Care Med, 2003,168(11) :1277-1292. 被引量:1
  • 9Tanaka M, Satoh M, Kawanami O, et al. A new bronchociber- scope for the study of diseases of very peripheral airways. Chest, 1984,85(5) :590-594. 被引量:1
  • 10Kikawada M, Ichinose Y, Miyamoto D, et al. Peripheral airway findings in chronic obstructive pulmonary disease using an ultra- thin bronchoscope. Eur Respir J, 2000,15(1) :105-108. 被引量:1

引证文献4

二级引证文献54

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部