金属对金属人工髋关节边缘接触效应 被引量:4

Effect of edge contact on metal-on-metal hip joint replacements
下载PDF
导出
摘要 目的研究金属对金属人工髋关节不同行走姿态下的接触力学行为,特别是大幅度运动可能导致的边缘接触效应。方法建立球面共型接触的髋关节有限元模型,通过改变髋臼相对股骨头的倾斜状态和对股骨头施加恒定竖直方向载荷相结合,等效模拟分析人工髋关节不同行走姿态下的接触状态。结果髋臼相对股骨头倾角在小于约60°范围内增加时,其对应的最大接触压力呈下降趋势,接触面积有所增大;当倾角超过80°范围时,关节接触区域因靠近髋臼边缘,最大接触压力位置由初始接触点向髋臼倾斜方向移动了约6°~9°的位置,用以满足压力分布合力与外载荷的平衡,接触压力和分布范围有所增加。结论髋臼相对股骨头较大倾斜状态极易引起不同的边缘接触现象,置换人体髋关节产生的边缘接触问题需要引起临床外科和关节制造上的重视。 Objective To study the contact mechanics of metal-on-metal hip joint replacements under different walking conditions, particularly for the edge contact behavior between the cup and the head of the bearing due to large range movement. Methods The finite element model of a typical hip joint replacement with conformal spherical bearing was established to simulate the edge contact of hip joint replacement under the given cup inclination angles and vertical loads. Results The maximum contact pressure decreased slightly and the contact area increased when the cup inclination angles increased within the range of 0 - 60. When the cup inclination an- gles were more than 80, the corresponding contact area moved to the cup edge, the maximum contact pressure and distribution range increased, but the position of the maximum contact pressure moved from the initial contact point to the new point with 6~9 in the direction of the increasing cup inclination angle to balance the corresponding contact pressure distribution and the loaded human weight. Conclusions Different edge contact behavior of hip joint replacements tends to occur due to the bigger cup inclination angle. Therefore, much attention should be paid to the problem of edge contact for both the clinical surgery and the manufacturing of hip joint replacements.
出处 《医用生物力学》 EI CAS CSCD 2011年第6期508-513,共6页 Journal of Medical Biomechanics
基金 国家自然科学基金资助项目(10972165) 国家重点基础研究发展计划(973)项目(2011CB706601)
关键词 人工髋关节 金属对金属 边缘接触 有限元法 接触压力 载荷 生物力学 Hip joint replacement Metal-on-metal Edge contact Finite element method Contact pressure Loads Biomechanics
  • 相关文献

参考文献11

  • 1Wang FC,Jin ZM.Transient elastohydrodynamic lubrica-tion of hip joint implants[J].J Tribol-Trans ASME,2008,130(1):011007(1-11). 被引量:1
  • 2王成焘.天然与人工关节中的摩擦学问题[J].医用生物力学,2009,24(5):317-325. 被引量:19
  • 3WANG FC,Jin ZM.Elastohydrodynamic lubrication model-ling of artificial hip joints under steady-state conditions[J].J Tribol,2005,127(4):729-739. 被引量:1
  • 4Wang FC,Wang LS,Sun ML.Tribological modelling ofspherical bearing with complex spherical-base geometryand motion[J].Trib Design,WIT Trans Eng Sci,2010,66:3-15. 被引量:1
  • 5Angadji A,Royle M,Colling S,et al.Influence of orienta-tion on the wear performance of metal-on-metal hip re-placements[J].Proc Inst Eng H,2009,223(4):449-457. 被引量:1
  • 6Liu H,Ellison PJ,Xu H,et al.Coupling of dynamics andcontact mechanics of artificial hip joints in a pendulum mod-el[J].Proc Inst Mech Eng H,2010,224(8):989-1003. 被引量:1
  • 7Besong AA,Lee R,Farrar R,et al.Contact mechanics ofa novel metal-on-metal total hip replacement[J].Proc InstMech Eng H,2001,215(6):543-548. 被引量:1
  • 8Wang FC,Zhao SX,Félix Quin-onez A,et al.Nonspherici-ty of bearing geometry and lubrication in hip joint replace-ments[J].J Tribol-Trans ASME,2009,131(3):031201. 被引量:1
  • 9Amstutz HC,Campell P,Kossovsky N,et al.Mechanismand clinical significance of wear debris-induced osteolysis[J].Clin Orthop Related Res,1992,276:7-18. 被引量:1
  • 10Hart AJ,Buddhev P,Winship P,et al.Cup inclination an-gle of greater than 50 degrees increases whole blood con-centrations of cobalt and chromium ions after metal-on-met-al hip resurfacing[J].Hip Int,2008,18(3):212-219. 被引量:1

二级参考文献5

共引文献18

同被引文献42

  • 1闫玉华,段湘慧.人工关节的研究现状和发展趋势[J].生物骨科材料与临床研究,2004,1(4):39-43. 被引量:22
  • 2倪小兵.人工髋关节设计现状与展望[J].生物骨科材料与临床研究,2004,1(2):55-56. 被引量:3
  • 3Scholes SC,Unsworth A.The tribology of metal-on-metaltotal hip replacements[J].Proc Inst Mech Eng H,2006,220(2):183-194. 被引量:1
  • 4Liu H,Ellison PJ,Xu H,et al.Coupling of dynamics andcontact mechanics of artificial hip joints in a pendulum mod-el[J].Proc Inst Mech Eng H,2010,224(8):989-1003. 被引量:1
  • 5Learmonth ID.Biocompatibility:A biomechanical and bio-logical concept in total hip replacement[J].Surgeon,2003,1(1):1-8. 被引量:1
  • 6Spears IR,Pfleiderer M,Schneider E,et al.Interfacialconditions between a press-fit acetabular cup and boneduring daily activities:Implications for achieving bone in-growth[J].J Biomech,2000,33(11):1471-1477. 被引量:1
  • 7Kassi JP,Heller MO,Stoeckle U,et al.Stair climbing ismore critical than walking in pre-clinical assessment of pri-mary stability in cementless THA in vitro[J].J Biomech,2005,38(5):1143-1154. 被引量:1
  • 8Besong AA,Lee R,Farrar R,et al.Contact mechanics ofa novel metal-on-metal total hip replacement[J].Proc InstMech Eng H,2001,215(6):543-548. 被引量:1
  • 9Wang FC,Wang LS,Sun ML.Tribological modelling ofspherical bearing with complex spherical-base geometryand motion[J].Tribol Design,WIT Trans Eng Sci,2010,66:3-15. 被引量:1
  • 10Anderson AE, Ellis BJ, Maas SA, et al. Effects of ideal- ized joint geometry on finite element predictions of cartilage contact stresses in the hip [J]. J Biomech, 2010, 43(7) : 1351-1357. 被引量:1

引证文献4

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部