摘要
目的了解Icare回弹式眼压计的临床实用性。方法选取受试者112例224只眼,先接受非接触眼压计测量,然后使用Icare回弹式眼压计测量眼压。结果Icare回弹式眼压计测量的右眼压值(15.4±3.4)mmHg,左眼压值(15.6±3.2)mmHg,非接触眼压计测量的右眼压值(13.5±2.7)mmHg,左眼压值(13.6±2.6)mmHg,两组比较差异有统计学意义(t分别为4.384,4.535;P〈0.01)。结论在正常眼压范围内,Icare回弹式眼压计测量结果高于非接触眼压计测量结果,Icare回弹式眼压计与非接触眼压计有很好的一致性。
Objective To discuss the clinical performance of ICARE rebound tonometer. Methods Intraocular pressure measurements were performed on 112 cases (224 eyes ) with non-contact tonometer and ICARE rebound tonometer respectively. Results The intraocular pressure of right eyes which measured by ICARE rebound tonometer was (15.4 ±3.4)mm Hg , and the left eyes' was (15.6 ±3.2)mm Hg on average , but they were (13.5 ±2.7)mm Fig and (13.6 ±2.6)rmn Fig respectively measured by non-contact tonometer There was significant difference between the two tonometer(t =4. 384,4. 535 ;P 〈 0. 01 ). Conclusions IOP measured by ICARE rebound tonometer is higher than that of non-contact tonometer when IOP is in the normal range. ICARE rebound and non-contact tonometer have a good consistency.
出处
《中华现代护理杂志》
2011年第34期4131-4133,共3页
Chinese Journal of Modern Nursing
关键词
眼内压
眼压测量
ICARE回弹式眼压计
非接触眼压计
Intraocular pressure
Intraocular pressure measurement
Icare rebound tonometer
Non-contact tonometer