期刊文献+

ANALYSIS OF AN OPEN-ENDED COAXIAL LINE SENSOR FOR BILAYERED BIOLOGICAL DIELECTRICS MEASUREMENT

ANALYSIS OF AN OPEN-ENDEDCOAXIAL LINE SENSOR FORBILAYERED BIOLOGICALDIELECTRICS MEASUREMENT
下载PDF
导出
摘要 This paper presents an analysis of an open-ended coaxial line sensor usedfor the dielectric propertty measurement or the bilayered biological tissues in vivo withwhich an analytical expression or sensor's fringing capacitances is derived. According tothis formula, an euquivalent circuit model of the open-ended coaxial line sensor is builtand a measurement method is designed for determining the dielectric properties or thebilayered materials. The theoretical results are compared with the experiment results,and it is shown that there is a good agreement between them. ANALYSISOFANOPEN-ENDEDCOAXIALLINESENSORFORBILAYEREDBIOLOGICALDIELECTRICSMEASUREMENTANALYSISOFANOPEN-ENDEDCOAXIALLINESENSORFOR...
出处 《Chinese Journal of Biomedical Engineering(English Edition)》 1994年第3期109-120,共12页 中国生物医学工程学报(英文版)
关键词 DIELECTRIC PROPERTY in VIVO MEASUREMENT dielectric property in vivo measurement