动态调强放射治疗滑窗技术的剂量和机械稳定性评价 被引量:4

Dosimetric and Mechanical Accuracy Estimation of Sliding Window in Dynamic Intensity-modulated Radiotherapy
下载PDF
导出
摘要 目的利用二维电离室矩阵和动态治疗日志文件,评价动态多叶准直器(MLC)滑窗技术的剂量学稳定性和机械精度。方法基于滑窗技术设计5种动态射野A、B、C、D、E,重复按照5种射野对二维电离室矩阵MatriXX进行照射并分别记录其动态日志文件,对12次的测量数据进行统计分析。结果各相应照射野等中心点剂量重复性分别为±1.49%、±1.12%、±1.12%、±1.07%、±1.07%,最大相对标准偏差为±1.47,γ分析通过率(3%/3mm)均为100%;动态治疗日志文件显示所有叶片定位误差均<0.05cm,定位误差的标准差均值分别为0.0365、0.038、0.038、0.035、0.026cm。结论本实验中动态MLC剂量学稳定性和机械精度满足治疗要求,本方法简单方便,可作为常规MLC质量保证程序的补充。 Objective We estimated dosimetric and mechanical accuracy of sliding window in dynamic intensity-modulated radiotherapy with a 2D diode array and dynalog file. Methods Five kinds of MLC format files based on sliding window technique were designed. 2D array of MatriXX was radiated in the linac and dynalog files were generated twelve times. Measured data was analyzed. Results All the repeatability errors of isocenter point dose of every fields were ± 1.49%,± 1.12%, ± 1.12%, ± 1.07%,± 1.07%, respectively. The maximum of relative standard deviations were ±1.47. All of γ- analyze pass rates were 100%. Dynalog file indicated that all leaf deviations were less than 0.5mm. The averages of errors RMS were 0.0365, 0.038, 0.038, 0.035, 0.026cm, respectively. Conclusion The result of the present work reveals the ideal functioning of the linac for dynamic dose delivery. This method is simple and convenient, which can be introduced into routine QA procedure for MLC.
出处 《中国医疗设备》 2011年第10期21-24,共4页 China Medical Devices
基金 广东省科技计划项目(2010B031600146) 广州医学院博士留学回国人员基金项目(2010-1-11)
关键词 直线加速器 多叶准直器 滑窗技术 剂量仪 调强放疗 linear accelerator multi-leaf collimator sliding window technique dosimeter intensitymodulated radiotherapy therapty
  • 相关文献

参考文献14

  • 1Ezzell G A,Galvin J M, Low D,et al.Guidance document on delivery,treatment planning, and clinical implementation of IMRT:report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee[J].Med.Phys,2003, 30:2089-2115. 被引量:1
  • 2Palta J, Kim S, Li J,et al.Tolerance limits and action levels for planning and delivery of IMRT [J]. Med. Phys,2003,30:1395. 被引量:1
  • 3曾自力.医用加速器多叶准直器的质量保证和质量控制[J].中国辐射卫生,2008,17(4):449-450. 被引量:6
  • 4S.S.Sivakumar,K.Krishna Murthy,C.A.Davis, R. Ravichandran,et al.Clinical implementation of dynamic intensity-modulated radiotherapy:Dosimetric aspects and initial experience[J].Med. Phys,2008,33:64-71. 被引量:1
  • 5Anup K Bhardwaj,TS Kehwar,SK Chakarvarti,et al. Dosimetric and qualitative analysis of kinetic properties of millennium 80 multileaf collimator system for dynamic intensity modulated radiotherapy treatments[J].Journal of Cancer Research and Therapeutics.2007,(3):23-28. 被引量:1
  • 6Venencia CD,Besa P.Commissioning and quality assurance for intensity modulated radiotherapy with dynamic mnltileaf collimator:Experience of the Pontificia Universidad Cat o lica de Chile[J].ApplClin Med Phys,2004,(5):37-54. 被引量:1
  • 7Mohammad Mohammadi and Eva Bezak. Evaluation of" MLC leaf" positioning using a scanning liquid ionization chamber EPID[J].Phys.Med, Biol.2007,(52): 21-33. 被引量:1
  • 8M Okumura,Y Obata,K Shimomura,M Tamura andY Nishimura.The effect of gantry and collimator angles on leaf limited velocity and position in dynamic multileaf collimator intensity-modulated radiation therapy[J].Phys Med Biol, 2010,(55):3101 -3113. 被引量:1
  • 9Clarke MF,Budgell GJ.Use of amorphous silicon EPID for measuring MLC calibration at varying gantry angle[J].Phys Med Biol.2008,53:473-485. 被引量:1
  • 10Kumar MD,Thirumavalavan N,Kfishna DV,et al.QA of Intensity modulated beams using dynamic MLC log files[J]. Med Phys,2006,31:36-41. 被引量:1

二级参考文献20

共引文献33

同被引文献40

  • 1吴昊,蒋璠,韩树奎,包尚联.常见探测器的空间分辨率比较[J].中国医学物理学杂志,2011,28(2):2501-2504. 被引量:9
  • 2戴建荣,胡逸民,张红志,关莹,张可,王闯.针对患者调强放射治疗计划的剂量学验证[J].中华放射肿瘤学杂志,2004,13(3):229-233. 被引量:65
  • 3鞠忠建,王运来,马林,徐寿平,戴相昆,王连元.用二维电离室矩阵验证多叶准直器叶片到位精度[J].中华放射肿瘤学杂志,2006,15(4):335-338. 被引量:18
  • 4Woo MK,Nico A. Impact of multi-leaf collimator leaf positioning accuracy on intensity modulation radiation therapy quality assurance ion chamber measurements[J].Medical Physics,2005,(05):1440-1445. 被引量:1
  • 5Chui CS,spirou S,LoSaaso T. Testing of dynamic multi-leaf collimation[J].Medical Physics,1996,(05):635-641. 被引量:1
  • 6Simon TA,Kahler D,Simon WE,et a1. An MLC calibration method using a detector array[J].Medical Physics,2009,(10):4495-4503. 被引量:1
  • 7Varian Medica Systems. Varain user manual.Dynalng File Viewer reference guide[M].2002. 被引量:1
  • 8J W Lee,J H Part,J B Chung. Inverse verification of the dose distribution for intensity modulated radiation therapy patientspecific quality assurance using dynamic MLC log files[J].Journal of the Koream Physical Society,2009,(04):1649-1656. 被引量:1
  • 9Litzenberg DW,Moran JM,Fraass BA. Verification of dynamic and segmental IMRT delivery by dynamic log file analysis[J].J App Clin Med Phys,2002,(02):63-72. 被引量:1
  • 10Klein EE,Hanley J,Bayouth J,et a1. Task group 142 report:quality assurance of medical accelerators[J].Medical Physics,2009,(10):4197-4212.doi:10.1118/1.3190392. 被引量:1

引证文献4

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部