摘要
聚变反应速率是表征惯性约束聚变热核反应的重要参数,为测量聚变反应速率,研制了一套由闪烁体及鼻锥部分、光学系统和条纹相机组成的测量系统。在神光Ⅲ原型装置上利用新研制的聚变反应速率测量系统进行了聚变反应速率测量。在DT中子产额约为1010条件下,首次获得了聚变反应速率随时间的变化过程。对影响聚变反应速率测量的相关因素进行详细分析后表明,系统的时间分辨力优于30ps。
Fusion reaction rate is an important parameter for measuring compression during the bnplosion in incrHal confine ment fusion experiment. We have developed a system for fusion reaction history measurement with high temporal resolution. The system is composed of plastic scintillator and nose cone, optical system and streak camera. We have applied this system on the SGⅢ prototype for fusion reaction rate measuring, For the first time, fusion reaction rate history have been measured for deutcrium tritium filled targets with netrons yields about 10^10. We have anatomized possible influence factor during fusion reaction rate meas uring. It indicates that the instrument measures fusion reaction bang time at temporal resolutions as low as 30 ps.
出处
《强激光与粒子束》
EI
CAS
CSCD
北大核心
2011年第8期2197-2200,共4页
High Power Laser and Particle Beams
基金
国家自然科学基金项目(10805041)
关键词
惯性约束聚变
聚变反应速率
热核反应
条纹相机
中子
inertial confinement fusion
fusion reaction rate
thermonuclear fusion
streak camera
nelron