期刊文献+

透析器自动复用机对透析溶质清除效果及透析器复用次数的影响 被引量:1

Effect of auto dialyzer reprocessing system on the solute clearance and the reusing times of the dialyzer
下载PDF
导出
摘要 目的:探讨透析器自动复用机对透析溶质的清除效果及合适的透析器复用次数。方法:收治40例慢性肾衰竭行血液透析患者,利用联机清除率监测器(OCM)测定患者在透析器使用第1次及复用第3,6,9次时K t/V值变化及透析器血室容积(TCV)的数值。结果:与第1次使用时比较,透析器在复用第9次后,约90%的TCV下降≥20%(P<0.05)。结论:透析器复用至第9次时,TCV下降≥20%,故透析器复用不应用超过9次。 Objective: To investigate the clearance effect of auto dialyzer reprocessing system on the dialysis solute and ap- propriate reusing times of the dialyzer. Methods : The hemodialysis therapy was applied to 40 patients with chronic renal failure, the on - line clearance monitor I OCM } was used to mensurate the changes of Kt / V value and the numerical value of the total cell volume ( TCV } of the dialyzer respectively at the first time of using the dialyzer and at the 3nd, 6th and 9th time of reusing the dialyzer. Results : The TCV declined from 90% to 20% at the 9th time compared with the first use of the dialyzer ( P 〈 O. 05 }. Conclusion : The reusing time of the dialyzer should be less than 9 times.
出处 《齐鲁护理杂志(上旬刊)》 2011年第8期6-7,共2页 Journal of Qilu Nursing
基金 广西壮族自治区卫生厅计划课题(Z2008307)
关键词 联机尿素清除率 血室容积 透析器 复用 On -line urea clearance rate Total cell volume Dialyzer Reuse
  • 相关文献

参考文献8

二级参考文献40

共引文献42

同被引文献5

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部