摘要
采用气相色谱-质谱联用仪,建立了测定塑料制品中三(2-氯乙基)磷酸酯含量的方法。本方法快速准确、灵敏度高,方法的线性范围是2~80mg/L,最低检出限为1mg/kg,回收率大于85%,结果满意。
A gas chromatography-mass spectrometry method for the determination of tris(2-chloroethyl)phosophate in plastics was developed.The method is fast,accurate and sensitive.The range of linearity is 2~80mg/L,the minimum detection limit is 1mg/L and the recovery is better than 85%.
出处
《分析仪器》
CAS
2011年第4期46-48,共3页
Analytical Instrumentation