摘要
目的探讨本地区新生儿网织红细胞计数(Ret)结果的性别差异,同时与传统的方法学比较,评价两种方法结果的可靠性。方法应用Sysmex XE-2100全自动血液分析仪和手工法同时进行计数。结果仪器法:男性新生儿Ret为(3.69±1.22)%,女性新生儿Ret为(2.92±1.10)%,手工法:男性新生儿Ret为(3.18±1.58)%,女性新生儿Ret为(2.41±1.30)%。结论 (1)男性与女性新生儿Ret结果经统计学处理差异无统计学意义;(2)两种方法比较的结果差异无统计学意义,但仪器法的重复性要好于手工法。
Objective To discuss the difference of reticulocytosis counting(Ret) results between different genders in local newborn babies and compare this methodology with traditional methodology.The reliability of these two methodologies were evaluated.Methods We used Sysmex XE-2100 automatic hematology analyzer and manual method to count respectively.Results The instrumental method: male newborn Ret(%)3.69 ± 1.22,female newborn Ret(%)for 2.92 ± 1.10;Manual method: male newborn Ret(%)3.18 ± 1.58,female newborn Ret(%) for 2.41 ± 1.30.Conclusion 1.There is no significant differerence between the reticulocytosis counting results of male and female newborn babies(P〉0.01).2.There is no significant difference between the results of these two methods,but the repeatability of the instrumental method is better than that of the manual method.
出处
《检验医学与临床》
CAS
2011年第9期1043-1044,共2页
Laboratory Medicine and Clinic
关键词
网织红细胞计数
计数方法
性别差异
重复性
reticulocyte
methods of counting
gender difference
repeatability