Pentacam系统和A型超声测厚仪获得的角膜中央区域厚度的相关性研究

Comparison of Central Corneal Thickness Measured by Pentacam vs A-ultrasound Pachymetry
下载PDF
导出
摘要 [目的]评估Pentacam系统和A型超声角膜测厚仪获得的角膜中央区厚度数据的一致性.[方法]应用Pentacam系统,对219人,共436眼,进行检查,分别获取角膜瞳孔中央厚度(作为角膜中央厚度,CCT)、角膜顶点厚度(CAT)和角膜最薄处厚度(CTT)数据;然后应用A型超声角膜测厚仪检测角膜中央区域厚度, 并对两种检查的数据进行统计学分析.[结果]Pentacam系统获得的角膜中央厚度为(556.42±35.00) μm,角膜顶点厚度为(555.69±35.02) μm,角膜最薄处厚度为(552.41±34.96) μm;超声角膜测厚仪获得的角膜中央厚度为(551.18±35.23) μm.两种仪器获得的数据呈直线正相关(r值均大于0.926,P值均小于0.05).经检验显示,Pentacam系统的角膜中央厚度和角膜顶点厚度与A型超声角膜中央厚度的差异有显著性(P<0.05);Pentacam系统的角膜最薄处厚度与超声角膜中央厚度的差异无显著性(P>0.05).应用Bland-Altman散点图方法分析,A型超声角膜中央厚度值偏低,更接近于Pentacam获得的角膜最薄处厚度值.[结论]Pentacam系统与A型超声角膜测厚仪获得的角膜中央厚度数据高度相关;在角膜厚度的整体观察方面,Pentacam系统优于A型超声角膜测厚仪;A型超声角膜测厚仪的角膜中央厚度数据与Pentacam系统的角膜最薄处厚度一致性好,临床工作中可以相互替换. [Objective]To assess the consistency of the central corneal thickness measured by the Pentacam system and A ultrasound pachymetry. [Methods] Pentacam system was used to measure the corneal pupil center thickness(as the central corneal thickriess, CCT), the corneal apex thickness(CAT) and the corneal thinnest thiekness(CTT) in 436 eyes of 219 participators, and then A-ultrasound paehymetry was used to measure the centraI corneal thickness of them. The data obtained by two examination methods were analyzed statistically. [Results] By Pentacam system, the CCT was (556, 42±35.00)μm, and the CAT was (555.69±35.02) μm, and the CTT was (552.41±34.96μm). By A ultrasound pachymetry, the CCT was (551. 18±35.23) μm. There was linear correlation between the data by two methods(r〉0, 926, all P 〈0.05). Paired samples T-test showed that there was statistical significance between CCT and CAT by Pentacam system and CCT by A-ultrasound paehymetry( P 〈0.05), and there was no statistical difference between CTT by Pentacam system and CCT by A uhrasound pachymetry( P 〉0.05). The Bland-Altman plots showed that the CCT by Aultrasound paehymetry was lower and close to the CTT by Pentacam system. [Conclusion]The CCT by Pentacam system is correlated with that by A-ultrasound pachymetry, By the whole observation of corneal thick ness, Pentacam system is better than A ultrasound pachymetry, There is good concordance between CCT A ultrasound pachymetry and CTT by Pentacam system, so they can be replaced in clinical work each other.
出处 《医学临床研究》 CAS 2011年第3期391-394,共4页 Journal of Clinical Research
关键词 角膜 cornea
  • 相关文献

参考文献12

  • 1Gherghel D,Hosking SL,Mantry S,et al.Corneal pachymetry in normal and keratoconic eyes:Orbscan Ⅱ versus ultrasound[J].Cataract Refract Surg,2004,30:1272-1277. 被引量:1
  • 2Medeiros FA,Sample PA,Zangwill LN,et al.Corneal thickness as a risk factor for visual field loss in patients with preperimetric glaucomatous optic neuropathy[J].Am J Ophthalmol,2003,136:805-813. 被引量:1
  • 3Javaloy J,Vidal MT,Villada JR,et al.Comparison of four corneal pachymetry techniques in corneal refractive surgery[J].J Refract Surg,2004,20:29-34. 被引量:1
  • 4Gonzalez-Meijome JM,Cervino A,Yebra-Pimentel E,et al.Central and peripheral corneal thickness measurement with Orbscan Ⅱ and topographical ultrasound pachymetry[J].J Cataract Refract Surg,2003,29:125-132. 被引量:1
  • 5Marsich MW,Bullimore M.The repeatability of corneal thickness measures[J].Cornea,2000,19:792-795. 被引量:1
  • 6Giasson C,Forthomme D.Comparison of central corneal thickness measurements between optical and ultrasound pachometers[J].Optom Vis Sci,1992,69:236-241. 被引量:1
  • 7Herse P,Siu A.Short-term effects of proparacaine on human corneal thickness[J].Acta Ophthalmol,1992,70:740-744. 被引量:1
  • 8Asensio I,Rahhal SM,Alonso L,et al.Corneal thickness values before and after oxybuprocaine 0.4% eye drops[J].Cornea,2003,2:527-532. 被引量:1
  • 9Andrew K.C.Lam,Davie Chen.Effect of proparacaine on central corneal thickness values[J].Cornea,2007,26:55-38. 被引量:1
  • 10Yaniv Barkana,Yariv Gerber.Central corneal thickness measurement with the Pentacam Scheimpflug system,optical low-coherence reflectometry pachymeter,and ultrasound pachymetry[J].J Cataract Refract Surg,2005,31:1729-1735. 被引量:1

二级参考文献4

  • 1赵希刚,李坤成,李永忠.单次屏气电影法MRI评价左心室功能的初步研究[J].中国医学影像技术,2005,21(1):65-67. 被引量:12
  • 2Bland JM,Altman DG.Statistical methods for assessing agreement between two methods of clinical measurement.Lancet,1986,i:307-310. 被引量:1
  • 3Bland JM,Altman DG.Measuring agreement in method comparison studies.Statis tical Methodsin Medical Research,1999,8:135-160. 被引量:1
  • 4Najib Aziz,Michael RI,Sally SD,et al.Interpreting Method Comparison Studiesby Useofthe Bland-Altman Plot:Reflecting the Importance of Sample Sizeby Incorporating Confidence Limitsand Predefined Error Limitsinthe Graphic.Clinical Chemistry,2004,50 (11):2216-2222. 被引量:1

共引文献216

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部