摘要
验证自动进样器进样量的准确性,符合中国药品GMP认证的要求。我们采用Nova- Pak18 HPLC色谱柱,30% 甲醇加70% 水为流动相,检测波长为272 nm ,咖啡因为测定物质;根据样品量与所测得峰面积呈正比的关系。
Validation of the accuracy of the HPLC Autosampler's injection is necessary for Chinese drug GMP authentication. We use the chromatographic system consisting of Nova-Pak C18 HPLC column and mobile phase of 30% methanol and 70% water with a detection wavelength at 272 nm and caffeine as detective samples. According to the direct ratio between quantity of sample to peak area (response), the accuracy of autosampler's injection is validated.
出处
《苏州城建环保学院学报》
1999年第3期78-82,共5页
Journal of Suzhou Institute of Urban Construction and Environmental Protection