摘要
The loss of a high-finesse Fabry-Perot cavity was measured using frequency scan and cavity ring-down techniques.Different transverse modes were examined by sweeping the frequency of the laser across the corresponding resonances.In addition,cavity ring-down technique was used to measure the loss of the Fabry-Perot cavity and a simple model was employed to remove the influence of finite response time of the optical switch and the detection circuit.
The loss of a high-finesse Fabry-Perot cavity was measured using frequency scan and cavity ring-down techniques.Different transverse modes were examined by sweeping the frequency of the laser across the corresponding resonances.In addition,cavity ring-down technique was used to measure the loss of the Fabry-Perot cavity and a simple model was employed to remove the influence of finite response time of the optical switch and the detection circuit.
出处
《中国激光》
EI
CAS
CSCD
北大核心
2010年第12期3075-3080,共6页
Chinese Journal of Lasers
基金
Supported by Bureau of Comprehensive Planning,Chinese Academy of Sciences