置入软组织金标数目对射波刀系统影像定位精度影响 被引量:17

Impact of number of implanted fiducials on image target localization accuracy
原文传递
导出
摘要 目的 通过硅橡胶模型模拟实验探讨置入软组织内金标数目对射波刀系统影像定位精度的影响.方法 模型内层为刚体结构,使用Accuray公司设计的球方模型;实验中用5个金标进行摆位以确定参考位置.模型外层为模拟人体软组织的非刚体性结构,使用双组分加成型硅橡胶;实验中用8个金标按不同数目的 排列组合获取影像定位数据,比较其与参考位置在平移参量和旋转参量方面差值以此探讨追踪不同数目金标对影像定位精度的影响.结果 追踪4个金标时平均每个金标对平移和旋转参量误差下降的贡献最大,分别为0.086 mm和0.033°.1~4个金标的平移参量误差下降了0.343 mm,4~8个金标的下降了0.077 mm.3~4个金标旋转参量误差下降了0.131°,4~8个金标的下降了0.090°.结论 追踪4个金标时平均每个金标对误差降低的贡献最大.金标置入4个以下时每多置1个金标对影像定位精度明显增加,置入4个以上时再增加置入金标数目对定位精度无明显改善.置入4个金标是同时兼顾降低患者痛苦和增加定位精度的一个适宜选择. Objective To find a proper number of implanted fiducials in order to reconcile both more accurate image tacking and less pain for the patients.Methods The phantom was made of two parts of different materials.The inner part was a ball-cube phantom developed by Accuray Corporation, which could be regarded as a rigid body, with 5 fiducials inside using as the reference to align the whole phantom in the experiment.The outer part was made of additional silicone rubber with two components to simulate soft tissue as a non-rigid body, which was implanted with 8 fiducials inside.All combinations of different number of fiducials were tracked to aquire the target location information, which were then compared with the reference position that was set by the inner 5 fiducials to obtain the deviations of the translation and the rotation parameters.Thus the impact of the number of fiducials on image target localization accuracy could be analyzed easily.Results When 4 fiducials are used for tracking, the decline of translation and rotation errors for every fiducial is largest (0.086 mm and 0.033°).The translation error was decreased by 0.343 mm from 1 to 4 fiducials, while by only 0.077 mm from 4 to 8 fiducials;the same decline was observed for rotation errors (0.131° for 3 to 4 fiducials , and 0.09° for 4 to 8 fiducials , respectively) .Conclusions When 4 fiducials are used for tracking, the decline of errors for every fiducial is maximum.When one more fiducial is adding, the image target localization accuracy can be enhanced obviously from 1 to 4 fiducials, but not as obvious for more than 4 fiducials.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2010年第5期465-467,共3页 Chinese Journal of Radiation Oncology
关键词 放射疗法 射波刀 金标数目 靶位置精度 Radiotherapy,cyberknife Number of fiducials Target localization accuracy
  • 相关文献

参考文献10

  • 1Adler JR Jr,Murphy MJ,Chang SD,et al.Image-guided robotic radiosergery.Neurosurgery,1999,44:1299-1306. 被引量:1
  • 2Harada T,Shirato H,Ogura S,et al.Real-time tumor-tracking radiation therapy for lung carcinoma by the aid of insertion of a gold mark using bronchofiberscopy.Cancer,2002,95:1720-1727. 被引量:1
  • 3Welsh JS,Berta C,Borzillary S,et al.Fiducial markers implanted during prostate brachytherapy for guiding conformal external beam radiation therapy.Technol Cancer Res Treat,2004,3:359-364. 被引量:1
  • 4Olender D.Fiducials for target localization//Heilbrun MP,eds.Cyberknife radiosurgery:a practical guide.Sunnyvale.CA:Cyberknife Society,2003:80-94. 被引量:1
  • 5Reichner CA,Collins BT,Gagnon GJ,et al.The placement of gold fiducials for cyberknife stereotactic radiosurgery using a modified transbronchial needle aspiration technique.J Bronchology,2005,12:193 -195. 被引量:1
  • 6Fu D,Kuduvalli G,Maurer CR,et al.3D target localization using 2D local displacements of skeletal structures in orthogonal X-ray images for image-guided spinal radiosurgery.Int J Comp Assis Radiol Surg,2006,Suppl 1:198-200. 被引量:1
  • 7Wu X,Fu D,Alberto DLZ,et al.Patient alignment and target tracking in radiosurgery of soft-tissue tumors using combined fiducial and skeletal structures tracking techniques//Harold CU,John JK,James DL,et al,eds.Treating tumors that move with respiration.Berlin:Springer,2007:31-36. 被引量:1
  • 8Brown WT,Perman M,Wu X,et al.Image-guided robotic stereotatic radiosurgery for treatment of lung tumors//Mould RF,Bucholz RD,Gagnon GJ,et al,eds.Robotic Radiosurgery.CA:The Cyberknife Society Pres,2005:255-268. 被引量:1
  • 9Perman M,Bellairs EE,Wu X,et al.Cancer of the pancreas with special reference to epidemiology& radiosurgery//Mould RF,Bucholz RD,Gagnon GJ,et al,eds.Robotic Radiosurgery.CA:The Cyberknife Society Press,2005:301-314. 被引量:1
  • 10Fu D,Robert K,Wang B,et al.X-sight lung tracking system:a fiducial-less method for respiratory motion tracking//Harold CU,John JK,James DL,et al,eds.Treating tumors that move with respiration.Berlin:Springer,2007:265 -282. 被引量:1

同被引文献86

  • 1Olivier Morin,Amy Gillis,Josephine Chen,Michèle Aubin,M.Kara Bucci,Jean Pouliot,索京涛,陆嘉德.兆伏级锥形束CT:系统说明及IGRT临床应用介绍[J].中国癌症杂志,2006,16(6):513-524. 被引量:9
  • 2国家质量监督检验检疫总局.GB18871-2002电离辐射防护与辐射源安全基本标准[S].北京:中国标准出版社,2003. 被引量:34
  • 3Haigron P, Dillenseger JL, Luo L, et al. Image-guided therapy: evolution and breakthrough. IEEE Eng Med Biol Mag, 2010,29: 100-104. 被引量:1
  • 4DeMarco JJ, Cagnon CH, Cody DD, et al. Estimating radiation doses from muhidetector CT using Monte Carlo simulations:effects of different size voxelized patient models on magnitudes of organ and effective dose. Phys Med Biol ,2007 ,52 :2583-2597. 被引量:1
  • 5Wong JR, Grimm L, Uematsu M,et al. Image-guided radiotherapy for prostate cancer by CT-linear accelerator combination: prostate movements and dosimetric considerations. Int J Radiat Oncol Biol Phys ,2005,61:561-569. 被引量:1
  • 6Daz-Santos CN, Amaro BJA, Cardiel GA,et al. The safety and efficacy of robotic image-guided radiosurgery system treatment for intra-and extracranial lesions:a systematic review of the literature. Radiother Onco1,2008 ,89 :245-253. 被引量:1
  • 7Ozhasoglu C, Saw CB, Chen H, et al. Synchrony--cyberknife respiratory compensation technology. Med Dosim, 2008,33 : 117- 123. 被引量:1
  • 8Pepin EW, Wu H, Zhang Y, et al. Correlation and prediction uncertainties in the cyberknife synchrony respiratory tracking system. Med Phys,2011,38:40364044. 被引量:1
  • 9Accuray Inc.Products[EB/OL].(2008-04-03)[2012-06-10].http://www.accuray.com/Products/Cyberknife/index.aspx. 被引量:1
  • 10Murphy M J.Tracking moving organs in real time[J].Semin Radi-atoncol,2004,14(1):91-100. 被引量:1

引证文献17

二级引证文献67

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部