XPERT非接触性眼压计的临床应用 被引量:4

A clinical evaluation of XPERT TM NCT(Reichert)
原文传递
导出
摘要 对XPERT非接触性眼压计进行准确性研究,为该眼压计在临床和青光眼筛选的应用提供可靠依据。方法:应用XPERT非接触性眼压计和Goldmann压平眼压计测定了293眼(150例)的眼压,并对其眼压值进行统计学分析。结果:XPERT非接触性眼压计和Goldmann压平眼压计测量值呈密切相关(r=0.9702,P=0.0000)。两者测定的均数差=0.1553(P>0.05)。以Goldmann压平眼压计测定为标准,XPERT非接触性眼压计测定值的波动范围在±0.133kPa(1mmHg)内占57.7%,在±0.4kPa(3mmHg)内的占85%,在±0.655kPa(5mmHg)内的占96.6%。结论:经研究显示XPERT非接触性眼压计和Goldmann压平眼压计的测量值一致性和相关性较好,可以应用于临床和人群青光眼检查快速筛检眼压升高者。 To study the accuracy of XPERT TM Non contact tonometer and provide a reliable data for its using in glaucoma clinics and screening.Methods:Intraocular pressure(IOP)of 150 patients(293 eyes)was measured by XPERT TM NCT(XPERT)and R900 Goldmann applanation tonometer(GAT),whose results were compared.Results:Mean difference between two methods(G NCT)was 0 02kPa(P>0 05).The linear correlation between two methods showed correlation relationship r=0 9702,P<0 0000.The distribution of G NCT according to the value of GAT showed that 57 7% data fluctuated between±0 133kPa,85% between ±0 4kPa,96 6% between ±0 655kPa.When IOP≥2 8kPa that measured by GAT is a cutoff point of screening the sensitivity of XPERT was 90 3%,specificity was 94 1%.Conclusions:Statistical analysis of these data showed no difference and a very good correlation in the two methods of measurement.XPERT TM NCT can be used in glaucoma clinics and screening.
出处 《中国实用眼科杂志》 CSCD 1999年第2期103-104,共2页 Chinese Journal of Practical Ophthalmology
关键词 青光眼 眼压计 非接触性 XPERT intraocular pressure Goldmann applanation tonometer non contact tonometer glaucoma
  • 相关文献

参考文献4

同被引文献57

  • 1刘丽,周跃华.影响近视患者眼压测量值的多因素分析[J].眼科新进展,2006,26(2):133-136. 被引量:9
  • 2韩英军,梁天蔚,勇志鹏,张满红,李文芳,胡利华,王立华,陈建丽,王婕,孙挥宇.Tono-pen眼压计与Goldmann眼压计对穿透性小梁切除术后患者眼压测量的比较[J].国际眼科杂志,2006,6(6):1424-1426. 被引量:12
  • 3Pandav SS,Sharma A,Gupta A,et al.Reliability of ProTon and Goldmann applanation tonometers in normal and postkeratoplasty eyes[J].Ophthalmology,2002,109(5):979-984. 被引量:1
  • 4Garzoz HJ,Chung HS,Lang Y,et al.Intraocular pressure and photorefractive keratectomy a comparison of three different tonometers[J].Cornea,2001,20(1):33-36. 被引量:1
  • 5Duch S,Serra A,Castanera J,et al.Tonometry after laser in situ keratomileusis treatment[J].J Glaucoma,2001,10(4):261-265. 被引量:1
  • 6Kaufmann C,Bachmann LM,Thiel MA.Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis[J].Invest Ophthalmol Vis Sci,2003,44(9):3790-3794. 被引量:1
  • 7Munoz LA,Perez AL,Rahhal MS,et al.Assessment of applanation tonometry after hyperopic laser in situ keratomileusis[J].Cornea,2002,21(2):156-160. 被引量:1
  • 8Copt RP,Tomas R,Mermoud A.Corneal thickness and normal tension glaucoma[J].Arch Ophthalmol,1999,117(1):14-16. 被引量:1
  • 9Dekker PW,Robert YCA,Kanngiesser H,et al.Principles of contact lens tonometry[J].International Ophthalmology,1998,22(2):105-111. 被引量:1
  • 10Shah S,Chatterjee A,Mathai M,et al.Relationship between corneal thickness and messured intraocular pressure in a general opthalmology clinic[J].Ophthalmology,1999,106(11):2154-2160. 被引量:1

引证文献4

二级引证文献40

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部