两种 LiF(Mg,Cu,P)片状探测器剂量学特性的比较研究 被引量:4

Comparative Study of Main Dosimetric Characteristics Between the GR 200A and MCP N LiF (Mg, Cu, P) TLD Chips
下载PDF
导出
摘要 GR200A(本实验室产品)与MCPN(波兰产品)均为LiF(Mg,Cu,P)片状探测器,现将GR200A和MCPN在发光曲线、灵敏度、重复使用、抗潮湿性能、本底、探测阈、光敏等性能进行平行对比实验,其中GR-200A的灵敏度为MCPN的16倍;6次重复使用的结果GR200A灵敏度无变化,MCPN灵敏度下降了26%;抗潮湿性能GR200A更优于MCPN。在98%的相对湿度下贮存25d后,GR200A灵敏度和颜色均无明显变化,而MCPN由原来的白色转变为淡棕色。实验结果表明:GR200A剂量学特性超过了MCPN。 GR 200A and MCP N are the same kind of LiF(Mg, Cu, P) chip detectors. The main dosimetric characteristics of two LiF (Mg, Cu, P) chip TLDs (GR 200A made in Solid Dosimetric Detector & Method Laboratory, Beijing, and MCP N made in Poland) are compared, including glow curve structure, TL sensitivity, anti moisture, reproducibility, background, detection threshold and photosensitivity. Sensitivity of GR 200a is 1 6 times as much as that of MCP N. The sensitivity of GR 200A is stable after being used for six times, but the sensitivity of MCP N dropped 2 8%, and its colour turned from white into light brown. The sensitivity and colour of GR 200A didnt vary after being stored for 25 days at 98% RH, by comparison the sensitivity of MCP N dropped 14%, and its colour turned into light brown. It was found that the GR 200A TLD is for superior to the MCP N TLD in above mentioned characteristics.
机构地区 北京防化研究院
出处 《核电子学与探测技术》 CAS CSCD 北大核心 1999年第2期90-93,共4页 Nuclear Electronics & Detection Technology
关键词 氟化锂 探测器 剂量计 剂量测量 热释光探测器 LiF(Mg, Cu, P) dosimeter Dosimeter Dosimetry Thermoluminescence detector