期刊文献+

湿度对便携式傅立叶红外快速检测气态污染物的影响

Effect of Humidity on the Fast Analysis of Gaseous Pollutant in Environment by Handy Transform Fourier Infrared Spectrometer
下载PDF
导出
摘要 用便携式傅立叶红外对不同湿度条件下的NO2、SO2、CS2和NO进行快速检测,探讨了气体中H2O含量对该方法的影响,试验结果对突发性污染事故的应急监测具有指导意义。 The fast analysis of NO2,NO,CS2 and SO2 in different humidity condition by handy fourier transform infrared spectrometer was studied and the effect of the water content contained in gas on the analysis method was discussed.The results can be used as data support for the fast analysis method of gaseous pollutant by handy fourier transform infrared spectrometer in emergency monitoring in sudden environmental pollution accident.
出处 《中国环境监测》 CAS CSCD 北大核心 2010年第2期20-22,共3页 Environmental Monitoring in China
关键词 红外 应急 湿度 气态污染物 Infrared Emergency Humidity Gaseous pollutant
  • 相关文献

参考文献5

二级参考文献23

  • 1林均挺.中国红外产品指南[M].北京:电子工业出版社,-.. 被引量:1
  • 2吴谨光.近代傅立叶变换红外光谱技术及应用:上卷[M].北京:科学技术出版社,1994.451—458. 被引量:1
  • 3布鲁克光谱仪器公司.[A]..傅立叶红外谱仪技术及庆用论文集[C].,2002.. 被引量:1
  • 4Hanst P, L. Pollution: Trace gas analysis ,in Fourier Transform Infrared Spectroscopy Application to Chemical systems[J]. New York: AcademicPress. 1979,2: 79. 被引量:1
  • 5Herffet W. F. Appl. Opt[J]. 1982,21:635. 被引量:1
  • 6Herget W. F. and Brasher J. D. Appl. Opt[J].1979,18:3404. 被引量:1
  • 7Herget W. F. Am. Lab[J]. 1982,14-72. 被引量:1
  • 8Herget W. F. and Brasher J. D. Appl. Opt[J].1980,19: 508. 被引量:1
  • 9Wang Jude, Bian Haiyan, Chen Zouru. Spectosc.Lett [J]. 1988,21:935. 被引量:1
  • 10Kimberly S. IS, Kevin IS. IS, Laura K. H, et al.Motor Vehicle Fleet Emission by OP-FTIR.Environ. Sci. Technol[J]. 2000, 34(5): 897-899. 被引量:1