7种不同设计的人工晶状体边缘散射光的对比观察 被引量:1

Comparison of Stray-light Effects of 7 Intraocular Lens Edge Designs
下载PDF
导出
摘要 目的观察和比较7种不同设计方式的人工晶状体(IOL)光学区边缘不同部位产生散射光的差异。方法用单色激光分别对三片式组和单片式组IOL的上缘和袢与光学区结合部进行不同角度的照射。用数码相机拍摄投射到视网膜平面上的光线图像,比较图像差异。结果在对上缘照射时,两组IOL因边缘形态不同而产生了图像差异;在对袢与光学区结合部照射时,三片式组IOL出现不同于单片式组的线状光影或中心旁致密光斑。结论IOL的边缘形态和三片式IOL袢与光学区结合部的不规则结构可能是引起IOL眼患者术后眩光的重要因素。 Objective To compare the stray-light effects caused by different parts of optic edge of 7 intraocular lens (IOL)edge designs. Methods Monochromatic laser beam was used to illuminate the upper edge and complex of three-piece and single-piece IOLs at several angles of incidence. Light images produced in the retinal plane were photographed with a digital camera. The differences in the light images were compared between three-piece and single-piece IOLs. Results There were differences in light images between 2 groups,which may be due to different shapes of edge design when the upper edge of IOL was illuminated. Three-piece IOL exhibited line images or dense patch, which differed with single-piece IOL when the complex of IOL was illuminated. Conclusion Both edge shape and irregular structure in three-piece IOL complex may be important factors for postoperative glare in pseudophakic patients.
出处 《中国医科大学学报》 CAS CSCD 北大核心 2010年第3期184-186,共3页 Journal of China Medical University
基金 辽宁省教育厅科学研究计划项目(2008225005)
关键词 IOL 结合部 散射光 术后眩光 intraocular len complex stray-light postoperative glare
  • 相关文献

参考文献9

  • 1张振平主编..人工晶状体屈光手术学[M].北京:人民卫生出版社,2009:598.
  • 2Mamalis N, Brubaker J, Davis D, et al. Complications of foldable intraocular lenses requiring explantation or secondary intervention-- 2007 survey update[J]. J Cataract Refract Surg,2008,34(9) : 1584- 1591. 被引量:1
  • 3RJ Landry, IK Ilev, TJ Pfefer, et al. Characterizing re?ections from intraocular lens implants [ J ]. Eye, 2007,21 ( 8 ) : 1083-1086. 被引量:1
  • 4Van Den Berg TJ,Van Rijn LJ,Michael R,et al.Straylight effects with aging and lens extraction [ J ].Am J Ophthalmol, 2007,144 ( 3 ) : 358-363. 被引量:1
  • 5Franchini A, Gallarati BZ,Vaccari E. Analysis of stray-light effects related to intraocular lens edge design [J]. J Cataract Refract Surg, 2004,30( 7 ) : 1531-1536. 被引量:1
  • 6Hofmann T,Zuberbuhler B,Cervino A,et al. Retinal straylight and complaint scores 18 months after implantation of the AcrySof monofocal and ReSTOR diffractive intraocular lenses [J]. J Refract Surg, 2009,25 ( 6 ) : 485 -492. 被引量:1
  • 7Holladay JT, Lang A, Portney V. Analysis of edge glare phenomenon in intraocular lens edge designs[J]. J Cataract Refract Surg, 1999,25 (6) :748-752. 被引量:1
  • 8Erie JC, Bandhauer MH. Intraocular lens surfaces and their relationship to postoperative glare [ J ]. J Cataract Refract Surg, 2003,29 (2) : 336-341. 被引量:1
  • 9ChilesA R.IOL 材料.ASIA-PACIFIC EyeWorld,2009,5(3):16-19. 被引量:1

同被引文献27

  • 1Boxer Wachler BS,Scruggs RT,Yuen LH,et al.Comparison of the Visian ICL and Verisyse phakic intraocular lenses for myopia from 6.00 to 20.00 diopters.J Refract Surg,2009,25:765-770. 被引量:1
  • 2Kamiya K,Shimizu K,lgarashi A,et al.Four-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia.Arch Ophthalmol,2009,127:845-850. 被引量:1
  • 3Strampelli B.Supportabilita di lenti ariliche in camera anteriore nella afachia.O nei vizi di refrazione.Annali di ottomologia o clinica oculistica,Parma,1954,80:75-82. 被引量:1
  • 4Kohnen T,Baumeister M,Magdowski G.Scanning electronmicroscropic characteristics of phakic intraocular lense.Ophthalmology,2000,107:934-939. 被引量:1
  • 5Risto JU,Esko A,Nida HS.Implantable contact lens for high myopia.J Cataract Refract Surg,2002,28:29-36. 被引量:1
  • 6Akcay L,Kaplan AT,Kandemir B,et al.Toric intraocular Collamer lens for high myopia astismatism after penetrating.J Cataract Refract Surg,2009,35:2161-2163. 被引量:1
  • 7Kamiya K,Shimizu K,Igarashi A,et al.Clinical outcomes and patient satisfaction after Vision Implantable Collamer Lens removal and phacoemulsification with intraocular lens implantation in eyes with induced cataract.Eye (Lond),2010,24:304-309. 被引量:1
  • 8Koivula A,Taube M,Zetterstr(o)m C.Phakic refractive lens:two-year results.J Refract Surg,2008,24:507-515. 被引量:1
  • 9Seo JH,Kim MK,Wee WR,et al.Effects of white-to-white diameter and anterior chamber depth on implantable collamer lens vault and visual outcome.J Refract surg,2009,25:730-738. 被引量:1
  • 10Verde CM,Teus MA,Arranz-Marquez E,et al.Medeanium posterior chamber phakic refractive lens to correct high myopia.J Refract Surg,2007,23:900-904. 被引量:1

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部